Yang Kai, Kwan Alexander L C, Boone John M
Department of Radiology, University of California, Davis Medical Center, 4860 Y Street, Suite 3100 Ellison Building, Sacramento, California 95817, USA.
Med Phys. 2007 Jun;34(6):2059-69. doi: 10.1118/1.2737263.
Computer simulation methods were used to evaluate the spatial resolution properties of a dedicated cone-beam breast CT system. X-ray projection data of a 70 microm nickel-chromium wire were simulated. The modulation transfer function (MTF) was calculated from the reconstructed axial images at different radial positions from the isocenter to study the spatial dependency of the spatial resolution of the breast CT scanner. The MTF was also calculated in both the radial and azimuthal directions. Subcomponents of the cone beam CT system that affect the MTF were modeled in the computer simulation in a serial manner, including the x-ray focal spot distribution, gantry rotation under the condition of continuous fluoroscopy, detector lag, and detector spatial resolution. Comparison between the computer simulated and physically measured MTF values demonstrates reasonable accuracy in the simulation process, with a small systematic difference (approximately 9.5 +/- 6.4% difference, due to unavoidable uncertainties from physical measurement and system calibration). The intrinsic resolution in the radial direction determined by simulation was about 2.0 mm(-1) uniformly through the field of view. The intrinsic resolution in the azimuthal direction degrades from 2.0 mm(-1) at the isocenter to 1.0 mm(-1) at the periphery with 76.9 mm from the isocenter. The results elucidate the intrinsic spatial resolution properties of the prototype breast CT system, and suggest ways in which spatial resolution can be improved with system modification.
采用计算机模拟方法评估专用锥形束乳腺CT系统的空间分辨率特性。模拟了70微米镍铬丝的X射线投影数据。从距等中心不同径向位置处重建的轴向图像计算调制传递函数(MTF),以研究乳腺CT扫描仪空间分辨率的空间依赖性。还在径向和方位方向上计算了MTF。在计算机模拟中以串行方式对影响MTF的锥形束CT系统子部件进行建模,包括X射线焦点分布、连续透视条件下的机架旋转、探测器滞后和探测器空间分辨率。计算机模拟与物理测量的MTF值之间的比较表明模拟过程具有合理的准确性,存在较小的系统差异(由于物理测量和系统校准不可避免的不确定性,差异约为9.5 +/- 6.4%)。通过模拟确定的径向固有分辨率在整个视野范围内均匀约为2.0 mm(-1)。方位方向的固有分辨率从等中心处的2.0 mm(-1)下降到距等中心76.9 mm处周边的1.0 mm(-1)。结果阐明了原型乳腺CT系统的固有空间分辨率特性,并提出了通过系统改进提高空间分辨率的方法。