Suppr超能文献

使用整体柱测定食品样品中咖啡因的高通量高效液相色谱法的开发与验证

Development and validation of a high-throughput high-performance liquid chromatographic assay for the determination of caffeine in food samples using a monolithic column.

作者信息

Tzanavaras Paraskevas D, Themelis Demetrius G

机构信息

Quality Control Department, Cosmopharm Ltd., P.O. Box 42, Korinthos 20100, Greece.

出版信息

Anal Chim Acta. 2007 Jan 2;581(1):89-94. doi: 10.1016/j.aca.2006.07.081. Epub 2006 Aug 5.

Abstract

The present study reports the development and validation of a high-throughput high-performance liquid chromatographic (HPLC) assay for the determination of caffeine in food samples. The analyte was separated rapidly from sample matrix using a short monolithic column (50 mm x 4.6 mm i.d.). The flow rate was 3.0 mL min(-1), while the mobile phase consisted of ACN/water (10:90, v/v). Caffeine was detected directly at 274n m. Under the optimal HPLC conditions, the sampling rate was 60 h(-1). The assay was validated for linearity, LOD and LOQ, precision, selectivity and ruggedness. The case of external calibration versus standard addition for the analysis of real samples was also examined. The proposed assay was applied to the analysis of beverages and coffee samples.

摘要

本研究报告了一种用于测定食品样品中咖啡因的高通量高效液相色谱(HPLC)分析方法的开发与验证。使用短整体柱(50 mm×4.6 mm内径)将分析物与样品基质快速分离。流速为3.0 mL min⁻¹,流动相由乙腈/水(10:90,v/v)组成。咖啡因在274 nm处直接检测。在最佳HPLC条件下,进样速率为60 h⁻¹。该分析方法针对线性、检测限和定量限、精密度、选择性和耐用性进行了验证。还考察了实际样品分析中外部校准与标准加入法的情况。所提出的分析方法应用于饮料和咖啡样品的分析。