Suppr超能文献

[人工晶状体的计算]

[Calculations for intraocular lenses].

作者信息

Haicl P, Boguszaková J, Havránek R, Síblová O

机构信息

Katedra oftalmologie ILF, Praha.

出版信息

Cesk Oftalmol. 1992 Jan;48(1):32-6.

PMID:1739995
Abstract

The authors deal with the postoperative dioptric deviation from the calculated optic potency of the emetropic intraocular lens, using formulae SRK II, SRK/T and Holladay's formula. Within the range +/- 1.5 D the deviation varied in 84% of the operated eyes where the intraocular lens was assessed according to formula SRK II; in 76% when formula SRK/T was used and in 74% when Holladay's formula was used. The authors give also an account of possible sources of errors during measurement, calculation and evaluation of results.

摘要

作者使用SRK II公式、SRK/T公式和霍拉迪公式,探讨了正视眼人工晶状体计算光学屈光度的术后屈光偏差。在±1.5 D范围内,根据SRK II公式评估人工晶状体时,84%的术眼出现偏差;使用SRK/T公式时为76%,使用霍拉迪公式时为74%。作者还阐述了测量、计算和结果评估过程中可能的误差来源。