Effect of exposure time on curing efficiency of polymerizing units equipped with light-emitting diodes.

作者信息

David José R, Gomes Osnara M, Gomes João C, Loguercio Alessandro D, Reis Alessandra

机构信息

Department of Dentistry, School of Dentistry, University of Ponta Grossa, Campus Uvaranas, Brazil.

出版信息

J Oral Sci. 2007 Mar;49(1):19-24. doi: 10.2334/josnusd.49.19.

Abstract

A study was conducted to evaluate the top and bottom hardness of two composites cured using polymerizing units equipped with light-emitting diodes [LED] (LEDemetron; Elipar FreeLight, Coltolux LED) and one quartz-tungsten halogen device [QTH] (Optilux 501) under different exposure times (20, 40 and 60 sec). A matrix mold 5 mm in diameter and 2 mm in depth was made to obtain five disc-shaped specimens for each experimental group. The specimens were cured by one of the light-curing units (LCUs) for 20, 40 or 60 sec, and the hardness was measured with a Vickers hardness-measuring instrument (50 g/30 sec). Data were subjected to three-way ANOVA and Tukey's test (alpha = 0.05). LED LCUs were as effective as the QTH device for curing both composites. A significant increase in the microhardness values were observed for all light LCUs when the exposure time was changed from 20 sec to 40 sec. The Z250 composite showed hardness values that were usually higher than those of the Charisma composite under similar experimental conditions. LED LCUs are as efficient for curing composites as the QTH device as long as an exposure time of 40 sec or higher is employed. An exposure time of 40 sec is required to provide composites with a uniform and high Knoop hardness when LED light-curing units are employed.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验