Differences in exhaled breath condensate pH measurements between samples obtained with two commercial devices.

作者信息

Prieto L, Ferrer A, Palop J, Domenech J, Llusar R, Rojas R

机构信息

Seccion de Alergologia, Hospital Universitario Dr. Peset, C/Gaspar Aguilar 90, 46017 Valencia, Spain.

出版信息

Respir Med. 2007 Aug;101(8):1715-20. doi: 10.1016/j.rmed.2007.02.023. Epub 2007 May 3.

Abstract

BACKGROUND

The objective of this study was to determine differences in EBC pH between samples obtained by RTube and EcoScreen, and to identify the effect of storage at -80 degrees C on the pH values.

METHODS

Twenty-three nonsmoking subjects with asthma or allergic rhinitis or without respiratory disease performed two sequential exhaled breath condensate (EBC) collections, using the RTube collection system and the EcoScreen condenser. EBC pH was measured immediately after collection and after storage at -80 degrees C for 8 weeks, without deaeration and repeated following deaeration with Argon.

RESULTS

In EBC samples without deaeration, the EcoScreen pH values were significantly higher than the RTube pH values (mean difference, 0.32; 95% CI, 0.21-0.44, P < 0.0001). In deaerated EBC samples, the EcoScreen pH values were also significantly higher than the RTube pH values (mean difference, 0.12; 95% CI, 0.01-0.25, P=0.04). For both EBC collection systems, storage for 8 weeks had a significant influence on pH of nondeaerated samples.

CONCLUSIONS

The present study demonstrates that EBC pH value is dependent on the collection device used and that the storage for 8 weeks had a significant influence on the pH of samples analyzed without deaeration.

摘要

背景

本研究的目的是确定使用RTube和EcoScreen采集的样本之间呼气末冷凝液(EBC)pH值的差异,并确定在-80℃储存对pH值的影响。

方法

23名患有哮喘或过敏性鼻炎或无呼吸系统疾病的非吸烟受试者,使用RTube采集系统和EcoScreen冷凝器进行了两次连续的呼气末冷凝液(EBC)采集。采集后立即测量EBC的pH值,并在-80℃储存8周后测量,不进行除气,并用氩气除气后重复测量。

结果

在未除气的EBC样本中,EcoScreen的pH值显著高于RTube的pH值(平均差异为0.32;95%可信区间为0.21-0.44,P<0.0001)。在除气的EBC样本中,EcoScreen的pH值也显著高于RTube的pH值(平均差异为0.12;95%可信区间为0.01-0.25,P=0.04)。对于两种EBC采集系统,8周的储存对未除气样本的pH值有显著影响。

结论

本研究表明,EBC的pH值取决于所使用的采集设备,并且8周的储存对未除气分析的样本的pH值有显著影响。