A biomechanical evaluation of a cannulated compressive screw for use in fractures of the scaphoid.

作者信息

Rankin G, Kuschner S H, Orlando C, McKellop H, Brien W W, Sherman R

机构信息

Department of Orthopaedics, University of Southern California School of Medicine, Los Angeles.

出版信息

J Hand Surg Am. 1991 Nov;16(6):1002-10. doi: 10.1016/s0363-5023(10)80059-0.

Abstract

The compressive force generated by a 3.5 mm ASIF cannulated cancellous screw with a 5 mm head was compared with that generated by a standard 3.5 mm ASIF screw (6 mm head), a 2.7 mm ASIF screw (5 mm head), and a Herbert screw. The screws were evaluated in the laboratory with the use of a custom-designed load washer (transducer) to the maximum compressive force generated by each screw until failure, either by thread stripping or by head migration into the specimen. Testing was done on paired cadaver scaphoids. To minimize the variability that occurs with human bone, and because of the cost and difficulty of obtaining human tissue specimens, a study was also done on polyurethane foam simulated bones. The 3.5 cannulated screw generated greater compressive forces than the Herbert screw but less compression than the 2.7 mm and 3.5 mm ASIF cortical screws. The 3.5 mm cannulated screw offers more rigid internal fixation for scaphoid fractures than the Herbert screw and gives the added advantage of placement over a guide wire.

摘要

将头部为5mm的3.5mm AO空心松质骨螺钉所产生的压缩力,与标准的3.5mm AO螺钉(头部为6mm)、2.7mm AO螺钉(头部为5mm)以及Herbert螺钉所产生的压缩力进行比较。在实验室中,使用定制设计的负载垫圈(传感器)对这些螺钉进行评估,直至每个螺钉产生最大压缩力直至失效,失效方式要么是螺纹滑脱,要么是头部移入标本中。测试在成对的尸体舟状骨上进行。为了尽量减少人类骨骼中出现的变异性,并且由于获取人体组织标本的成本和难度,还对聚氨酯泡沫模拟骨进行了研究。3.5mm空心螺钉产生的压缩力比Herbert螺钉大,但比2.7mm和3.5mm AO皮质螺钉产生的压缩力小。与Herbert螺钉相比,3.5mm空心螺钉为舟状骨骨折提供了更坚固的内固定,并且具有通过导丝置入的额外优势。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验