Abstract

At the Northern Illinois University Institute for Neutron Therapy at Fermilab, the clinical tissue-equivalent ionisation chamber response is measured every treatment day using a cesium source that was configured to match readings obtained at the National Bureau of Standards. Daily measurements are performed in air using the air-to-tissue dose conversion factors given in AAPM Report #7. The measured exposure calibration factors have been tabulated and graphed as a function of time from 1978 to present. For A-150 plastic ionisation chambers, these factors exhibit a sinusoidal variation with a period of approximately 1 y and amplitude of +/- 1%. This variation, attributable to the hygroscopic nature of A-150 plastic, is correlated with the relative humidity of the facility, and is greater than the humidity corrections for gas described in the literature. The data suggest that chamber calibration should be performed at least weekly to accommodate these variations.

摘要

在费米实验室的北伊利诺伊大学中子治疗研究所,每个治疗日都会使用一个铯源测量临床组织等效电离室响应,该铯源的配置使其读数与国家标准局获得的读数相匹配。每日测量在空气中进行,使用AAPM报告第7号中给出的空气与组织剂量转换因子。从1978年至今,已将测量的照射校准因子制成表格并绘制成时间的函数图。对于A - 150塑料电离室,这些因子呈现出近似1年周期的正弦变化,幅度为±1%。这种变化归因于A - 150塑料的吸湿性,与设施的相对湿度相关,且大于文献中描述的气体湿度校正。数据表明,应至少每周进行一次电离室校准以适应这些变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验