Suppr超能文献

五种肝素治疗床边监测设备的比较。

A comparison of five devices for the bedside monitoring of heparin therapy.

作者信息

O'Neill A I, McAllister C, Corke C F, Parkin J D

机构信息

Department of Haematology, Repatriation General Hospital, Heidelberg West, Victoria.

出版信息

Anaesth Intensive Care. 1991 Nov;19(4):592-6. doi: 10.1177/0310057X9101900423.

Abstract

Five instruments were tested for their capacity to monitor heparin therapy on whole blood at the bedside. The instruments were 512 Coagulation Monitor (Ciba-Corning), Thrombotrack (Nycomed), Automated Coagulation Timer (Hemotec), Hemochron-ACT and Hemochron-APTT (International Technidyne Corporation). Fifty subjects with various levels of heparinisation were tested on each instrument and were also assayed for antithrombin III, fibrinogen, haematocrit, platelet count and plasma heparin level. The results were compared with a reference APTT performed on the Automated Coagulation Laboratory 300R (Instrumentation Laboratories). The Hemochron-ACT correlated least well. The Hemotec and Thrombotrack were unsuitable in a clinical setting because of pipetting requirements, although the Thrombotrack did correlate well with the reference parameters. The 512 Coagulation Monitor was the simplest to use, but its maximum response corresponded to the midpoint of the reference APTT therapeutic range. The Hemochron-APTT was simple to use, had an adequate response range and correlated well with reference parameters.

摘要

对五种仪器进行了测试,以评估其在床边监测全血肝素治疗的能力。这些仪器分别是512凝血监测仪(汽巴-康宁公司)、血栓追踪仪(奈可明公司)、自动凝血计时器(赫莫泰克公司)、Hemochron-ACT和Hemochron-APTT(国际技术公司)。对50名肝素化程度不同的受试者在每种仪器上进行了测试,并对其抗凝血酶III、纤维蛋白原、血细胞比容、血小板计数和血浆肝素水平进行了检测。将结果与在自动凝血实验室300R(仪器实验室)上进行的参考活化部分凝血活酶时间(APTT)进行了比较。Hemochron-ACT的相关性最差。赫莫泰克和血栓追踪仪由于移液要求在临床环境中不适用,尽管血栓追踪仪与参考参数的相关性很好。512凝血监测仪使用最简单,但其最大反应对应于参考APTT治疗范围的中点。Hemochron-APTT使用简单,有足够的反应范围,并且与参考参数相关性良好。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验