Schmidt Gerwin P, Wintersperger Bernd, Graser Anno, Baur-Melnyk Andrea, Reiser Maximilian F, Schoenberg Stefan O
Institute of Clinical Radiology, University Hospitals Munich-Grosshadern, LMU, Munich, Germany.
Invest Radiol. 2007 Jun;42(6):449-59. doi: 10.1097/01.rli.0000262089.55618.18.
To analyze the impact of altered magnetic field properties on image quality and on potential artifacts when an established whole-body magnetic resonance imaging (WB-MRI) protocol at 1.5 Tesla (T) is migrated to 3 T.
Fifteen volunteers underwent noncontrast magnetic resonance imaging (MRI) on 32-channel whole body-scanners at 1.5 and 3 T with the use of parallel acquisition techniques (PAT). Coronal T1-weighted TSE- and short tau inversion recovery (STIR)-sequences at 4 body levels including sagittal imaging of the whole spine were performed. Additional axial HASTE-imaging of lung and abdomen, T1-/T2-weighted-TSE- and EPI-sequences of the brain and T2-weighted respiratory-triggered imaging of the liver was acquired. Both data sets were compared by 2 independent readers in respect to artifacts and image quality using a 5-point scale. Regions of pronounced artifacts were defined.
Overall image impression was both qualitatively rated as "good" at 1.5 and 3 T for T1-w-TSE- and STIR-imaging of the whole body and spine. At 1.5 T, significantly better quantitative values for overall image quality were found for WB-STIR, T2-w-TSE imaging of the liver and brain (Wilcoxon Mann-Whitney U Test; P < 0.05), overall rated as good at 3 T. Significantly higher dielectric effects at 3 T were affecting T1-w- and STIR-WB-MRI, and HASTE of the abdomen and better image homogeneity at 1.5 T was observed for T1-weighted-/STIR-WB-MRI and T1-w-TSE-imaging of the spine. Pulsation artifacts were significantly increased at 3 T for T1-w WB-MRI. Significantly higher susceptibility artifacts were found for GRE-sequences of the brain at 3 T. Motion artifacts, Gibbs-Ringing, and image distortion was not significantly different and showed slightly higher quantitative values at 3 T (except for HASTE imaging of the abdomen). Overall scan time was 45 minutes and 44 seconds at 1.5 T and 40 minutes and 28 seconds at 3 T at identical image resolution.
Three Tesla WB-MRI is feasible with good image quality comparable to 1.5 T. 3.0 T WB-MRI shows significantly more artifacts with a mild to moderate impact on image assessment. Therefore 1.5 T WB-MRI is the preferred image modality. Overall scan time at 3 T is reduced with the use of parallel imaging at a constant image resolution.
分析当将1.5特斯拉(T)的既定全身磁共振成像(WB-MRI)协议迁移至3T时,磁场特性改变对图像质量和潜在伪影的影响。
15名志愿者在1.5T和3T的32通道全身扫描仪上使用并行采集技术(PAT)进行非增强磁共振成像(MRI)。在4个身体层面进行冠状位T1加权快速自旋回波(TSE)和短tau反转恢复(STIR)序列扫描,包括整个脊柱的矢状位成像。还采集了肺部和腹部的额外轴向半傅里叶单次激发快速自旋回波(HASTE)成像、脑部的T1加权/T2加权TSE和回波平面成像(EPI)序列以及肝脏的T2加权呼吸触发成像。由2名独立阅片者使用5分制对两组数据集的伪影和图像质量进行比较。定义明显伪影的区域。
对于全身和脊柱的T1加权TSE和STIR成像,在1.5T和3T时总体图像印象在定性上均被评为“良好”。在1.5T时,全身STIR、肝脏和脑部的T2加权TSE成像的总体图像质量定量值显著更好(Wilcoxon Mann-Whitney U检验;P<0.05),在3T时总体评为良好。3T时显著更高的介电效应影响T1加权和STIR WB-MRI以及腹部的HASTE成像,并且在1.5T时观察到T1加权/STIR WB-MRI和脊柱的T1加权TSE成像具有更好的图像均匀性。对于T1加权WB-MRI,3T时搏动伪影显著增加。在3T时,脑部的梯度回波(GRE)序列发现显著更高的磁化率伪影。运动伪影、吉布斯环影和图像畸变无显著差异,在3T时显示出略高的定量值(腹部的HASTE成像除外)。在相同图像分辨率下,1.5T时的总扫描时间为45分钟44秒,3T时为40分钟28秒。
3T WB-MRI是可行的,图像质量良好,与1.5T相当。3.0T WB-MRI显示出明显更多的伪影,对图像评估有轻度至中度影响。因此,1.5T WB-MRI是首选的成像方式。在恒定图像分辨率下使用并行成像可减少3T时的总扫描时间。