Suppr超能文献

确定平衡电动液压全人工心脏所需的心房分流大小。

Determination of atrial shunt size needed to balance an electrohydraulic total artificial heart.

作者信息

Kinoshita M, Hansen C A, Khanwilkar P S, White K R, Olsen D B

机构信息

Artificial Heart Research Laboratory, University of Utah, Salt Lake City 84103-1414.

出版信息

ASAIO Trans. 1991 Jul-Sep;37(3):M264-5.

PMID:1751140
Abstract

An interatrial shunt (IAS) was developed and tested to balance a volumetrically coupled electrohydraulic total artificial heart (EHTAH). The IAS diameter was determined theoretically with the equation: [formula: see text] where D = orifice diameter, delta P = pressure drop across the orifice, rho = fluid density, Q = volume flow through the orifice, and C = empiric constant. The calculated orifice diameter was 2.9-5.3 mm to keep a 3-5 mmHg pressure difference between left and right atria, with 0.42-1.05 L/min of expected flow through the IAS. Three different sizes (3.4, 5.4, and 5.5 mm) of IAS were implanted and tested in four calves. The in vivo left-right pressure difference ranged from 3 to 8 mmHg, with 0.57-1.38 L/min of theoretically calculated flow through the IAS. At autopsy, none of the IAS showed macroscopic thrombus formation.

摘要

研发并测试了一种房间隔分流器(IAS),用于平衡容积耦合的电动液压全人工心脏(EHTAH)。IAS的直径通过以下公式理论确定:[公式:见原文],其中D = 孔口直径,ΔP = 孔口两端的压降,ρ = 流体密度,Q = 通过孔口的体积流量,C = 经验常数。计算得出的孔口直径为2.9 - 5.3毫米,以保持左右心房之间3 - 5毫米汞柱的压差,预计通过IAS的流量为0.42 - 1.05升/分钟。将三种不同尺寸(3.4、5.4和5.5毫米)的IAS植入四只小牛体内并进行测试。体内左右压差范围为3至8毫米汞柱,通过IAS的理论计算流量为0.57 - 1.38升/分钟。尸检时,所有IAS均未显示宏观血栓形成。