Bagley B, Bagley A, Henrie J, Froerer C, Brohamer J, Burkart J, Mortensen J D
CardioPulmonics, Salt Lake City, UT.
ASAIO Trans. 1991 Jul-Sep;37(3):M413-5.
The volume of O2 and CO2 transferred into/out circulating venous blood by various sized IVOX devices has been assessed by ex vivo and in vivo animal experiments and by reviewing the gas transfer data collected from the first 20 human ARDS patients studied during Phase I of the IVOX clinical trials. Data from these assessments indicate that oxygen transfer through the size 5 IVOX ranges from 15.4 to 18.0 cc/min; through the size 6 IVOX ranges from 22.8 to 35.5 cc/min; through the size 7 IVOX ranges from 28.5 to 66.7 cc/min; through the size 8 IVOX ranges from 34.9 to 66.3 cc/min; through the size 9 IVOX ranges from 45.9 to 115.7 cc/min; through the size 10 IVOX ranges from 52.9 to 133 cc/min. Quantitative carbon dioxide transfer through the various size IVOX devices closely follows the quantitative oxygen transfer achieved by IVOX. These quantitative data provide valuable information to clinicians considering using IVOX to augment the inadequate gas transfer existing in patients with acute, potentially reversible respiratory failure. They indicate IVOX transfers clinically significant quantities of O2 into and CO2 out of circulating venous blood in an intact subject without involving the subjects natural lungs.
通过体外和体内动物实验,以及回顾在IVOX临床试验第一阶段研究的首批20例人类急性呼吸窘迫综合征(ARDS)患者收集的气体传输数据,对各种尺寸的IVOX装置向/从循环静脉血中转移的氧气和二氧化碳量进行了评估。这些评估数据表明,通过5号IVOX装置的氧气转移量为15.4至18.0毫升/分钟;通过6号IVOX装置的氧气转移量为22.8至35.5毫升/分钟;通过7号IVOX装置的氧气转移量为28.5至66.7毫升/分钟;通过8号IVOX装置的氧气转移量为34.9至66.3毫升/分钟;通过9号IVOX装置的氧气转移量为45.9至115.7毫升/分钟;通过10号IVOX装置的氧气转移量为52.9至133毫升/分钟。通过各种尺寸IVOX装置的二氧化碳定量转移情况与IVOX实现的氧气定量转移情况密切相关。这些定量数据为考虑使用IVOX来改善急性、潜在可逆性呼吸衰竭患者存在的气体转移不足的临床医生提供了有价值的信息。它们表明,IVOX在不涉及受试者天然肺脏的情况下,能将临床上大量的氧气转移到循环静脉血中,并将二氧化碳从循环静脉血中转移出来。