Performance of an intravenous gas exchanger (IVOX) in a venovenous bypass circuit.

作者信息

Tao W, Zwischenberger J B, Nguyen T T, Tzouanakis A E, Matheis E J, Traber D L, Bidani A

机构信息

Department of Surgery, University of Texas Medical Branch, Galveston 77555-0528.

出版信息

Ann Thorac Surg. 1994 Jun;57(6):1484-90; discussion 1490-1. doi: 10.1016/0003-4975(94)90105-8.

Abstract

To analyze quantitatively the performance of the intravenacaval blood gas exchanger (IVOX), we developed a right atrium-pulmonary artery venovenous extracorporeal bypass circuit. Oxygen transfer and carbon dioxide removal were calculated at different blood flow rates, different hemoglobin levels, and during permissive hypercapnia. Oxygen transfer increased linearly with blood flow up to 41 mL/min. Likewise, O2 transfer increased linearly with hemoglobin levels up to 7.5 g/dL, but no further increases were achieved above this level. Carbon dioxide removal increased linearly as flow increased from 1.0 to 3.0 L/min but did not increase further for higher flows. Carbon dioxide removal was 45 mL/min at blood carbon dioxide tension of 42 mm Hg but increased to a maximum of 81 mL/min at a carbon dioxide tension of 90 mm Hg. We conclude that IVOX is a diffusion-limited device dependent on blood flow, hemoglobin content, and the gas pressure gradient across the membrane. Further engineering improvements are needed to improve the gas exchange performance of IVOX.