Development of an automatic control algorithm for the electrohydraulic total artificial heart without transducers.

作者信息

Kim H C, Khanwilkar P S, Crump K C, Smith K D, Bearnson G B, Olsen D B

机构信息

Institute for Biomedical Engineering, University of Utah, Salt Lake City.

出版信息

ASAIO Trans. 1991 Jul-Sep;37(3):M501-3.

PMID:1751252
Abstract

To achieve a reliable and simple implantable total artificial heart, the number of implanted transducers providing the physiologic information required for automatic control should be minimized. To address this need, a new automatic control algorithm, based on a transducerless electrohydraulic total artificial heart (EHTAH) system, is proposed. The current EHTAH physiologic control algorithm relies on two implanted pressure transducers. Without these transducers, the information required for automatic control must be extracted from the running motor's parameters. These parameters correlated with the differential pressure across the axial flow pump used to actuate the EHTAH. Changes in this differential were chosen as a cue for cardiac output control. This algorithm can be viewed as depending upon systemic vascular resistance determined by subtracting mean right atrial pressure (RAP) from mean aortic pressure (AoP) and dividing the result by total cardiac output (CO). The difference between mean AoP and mean RAP was confirmed to correlate with the differential hydraulic pressure across the energy converter during the left systolic phase. As an interim configuration, a single differential pressure transducer measuring the differential hydraulic pressure across the energy converter was tested on a Donovan mock circulation system. The resultant CO response shows good sensitivity according to changes in both preload and afterload.

摘要

相似文献

1
Development of an automatic control algorithm for the electrohydraulic total artificial heart without transducers.
ASAIO Trans. 1991 Jul-Sep;37(3):M501-3.
2
In vitro analysis of an atrial shunt in balancing an electrohydraulic total artificial heart.用于平衡电动液压全人工心脏的心房分流器的体外分析
ASAIO Trans. 1990 Jul-Sep;36(3):M254-7.
3
Right-left ventricular balance in implanted electrically powered artificial hearts.
ASAIO Trans. 1990 Jul-Sep;36(3):M287-90.
4
In vivo long-term evaluation of the Utah electrohydraulic total artificial heart.
ASAIO J. 1993 Jul-Sep;39(3):M373-80.
5
Development of a microcontroller-based automatic control system for the electrohydraulic total artificial heart.基于微控制器的电动液压全人工心脏自动控制系统的开发。
IEEE Trans Biomed Eng. 1997 Jan;44(1):77-89. doi: 10.1109/10.553715.
6
In vitro development of automatic control for the actively filled electrohydraulic heart.主动填充式电动液压心脏自动控制的体外研发
Artif Organs. 1988 Apr;12(2):152-62. doi: 10.1111/j.1525-1594.1988.tb02748.x.
7
Cardiac output regulation in the moving actuator total artificial heart without a compliance chamber.无顺应腔的可移动驱动型全人工心脏的心输出量调节
ASAIO J. 1992 Oct-Dec;38(4):846-50.
8
A compact and noise free electrohydraulic total artificial heart.
ASAIO J. 1993 Jul-Sep;39(3):M386-91.
9
A new automatic cardiac output control algorithm for moving actuator total artificial heart by motor current waveform analysis.一种通过电机电流波形分析对移动驱动器全人工心脏进行自动心输出量控制的新算法。
Int J Artif Organs. 1996 Mar;19(3):189-96.
10
Induced pulsation of a continuous-flow total artificial heart in a mock circulatory system.在模拟循环系统中对连续流全人工心脏进行脉动诱导。
J Heart Lung Transplant. 2010 May;29(5):568-73. doi: 10.1016/j.healun.2009.12.004. Epub 2010 Feb 13.