使用LiF:Mg,Ti热释光剂量计评估热中子剂量时读出参数的优化

Optimisation of the readout parameters when evaluating thermal neutron doses by TL dosimetry with LiF:Mg,Ti.

作者信息

German U, Weinstein M, Abraham A, Alfassi Z B

机构信息

Nuclear Research Center Negev, P.O. Box 9001, 84190 Beer Sheva, Israel.

出版信息

Radiat Prot Dosimetry. 2007;126(1-4):532-5. doi: 10.1093/rpd/ncm107. Epub 2007 May 18.

Abstract

The location of the glow peaks depends on the heating rate. It takes some time until the crystal reaches the heater temperature, and this time lag has a direct effect on the shift of peaks towards higher temperatures. Some information on the high-temperature peaks may be lost if the readout conditions (mainly length of time) are not properly chosen. Step heating profiles to a varying final temperature between 300 degrees C and 125 degrees C were used to study the time dependence of the collected information in the glow curves of (6)LiF:Mg,Ti crystals, and the minimal heating time for evaluation of thermal neutron doses was determined.

摘要

发光峰的位置取决于加热速率。晶体达到加热器温度需要一些时间,而这段时间延迟会直接影响峰向更高温度的偏移。如果读出条件(主要是时间长度)选择不当,高温峰的一些信息可能会丢失。采用逐步升温至300℃至125℃之间不同最终温度的加热曲线,研究了(6)LiF:Mg,Ti晶体发光曲线中收集信息的时间依赖性,并确定了评估热中子剂量的最小加热时间。