Lin Roger C, Li Yan, Tang Maolong, McLain Marcy, Rollins Andrew M, Izatt Joseph A, Huang David
Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.
Cornea. 2007 Jun;26(5):594-9. doi: 10.1097/ICO.0b013e31803c5535.
To use optical coherence tomography (OCT) to detect previous refractive surgery in donor corneas.
We constructed a tabletop OCT scanner operating at 1310-nm wavelength. Donor corneas at the Cleveland Eye Bank were scanned while sealed within the sterile container immersed in Optisol GS. OCT scanning was performed with 7.6-mm-long lines (512 axial scans) along 8 meridians. Anterior and posterior corneal surfaces were automatically mapped using image processing software that we developed. Curvature was computed from the best parabolic fit in the central 5-mm diameter. Layered analysis of the stromal reflectivity was also performed. Twenty-nine corneas from 19 donors were examined. Five had a history of laser in situ keratomileusis (LASIK).
The flap interfaces could not be visualized on slit-lamp or OCT images but were confirmed by histology. The death-to-scan time was 22.1 +/- 11.4 (SD) hours for normal corneas and 100.6 +/- 57.5 hours for LASIK corneas. The anterior surface power was 67.5 +/- 2.5 D in control corneas and 64.5 +/- 2.4 D in LASIK corneas (P = 0.023). There was no significance between the 2 groups in terms of posterior curvature and thickness parameters. The anterior/posterior reflectivity ratio in the central 4-mm diameter was significantly lower in post-LASIK corneas than in control (P < 0.05).
OCT provides thickness, topography, and reflectivity maps of donor corneas without taking them out of preservation medium and container. The anterior curvature and the anterior/posterior stromal reflectivity ratio may be useful for detecting previous LASIK.
使用光学相干断层扫描(OCT)检测供体角膜先前的屈光手术情况。
我们构建了一台工作波长为1310纳米的桌面式OCT扫描仪。对克利夫兰眼库的供体角膜进行扫描,角膜密封在浸于Optisol GS中的无菌容器内。沿着8条子午线进行7.6毫米长的线扫描(512次轴向扫描)。使用我们开发的图像处理软件自动绘制角膜前表面和后表面。通过对中央直径5毫米范围内的最佳抛物线拟合计算曲率。还对基质反射率进行分层分析。检查了来自19名供体的29只角膜。其中5只角膜有准分子原位角膜磨镶术(LASIK)病史。
在裂隙灯或OCT图像上无法看到瓣界面,但经组织学证实。正常角膜的死亡至扫描时间为22.1±11.4(标准差)小时,LASIK角膜为100.6±57.5小时。对照角膜的前表面屈光力为67.5±2.5D,LASIK角膜为64.5±2.4D(P = 0.023)。两组在角膜后曲率和厚度参数方面无显著差异。LASIK术后角膜中央直径4毫米范围内的前/后反射率比值显著低于对照组(P < 0.05)。
OCT无需将供体角膜从保存介质和容器中取出即可提供其厚度、地形图和反射率图。角膜前曲率以及前/后基质反射率比值可能有助于检测先前的LASIK手术。