Langner F, Löb S, Nolte R, Gressier V, Asselineau B, Lacoste V, Lebreton L
Physikalisch-Technische Bundesanstalt, Braunschweig, Germany.
Radiat Prot Dosimetry. 2007;126(1-4):145-50. doi: 10.1093/rpd/ncm031. Epub 2007 May 25.
The photon contribution to ambient dose equivalent in several wide-spectrum reference neutrons fields of the Institute for Radiological Protection and Nuclear Safety were measured using a Geiger-Müller counter. For the investigated fields, the ratio of photon to neutron ambient dose equivalent ranged between 0.03 and 0.20. The results show that the Geiger-Müller tube is a versatile instrument for dosimetry in mixed photon-neutron fields if sufficient information for the calculation of corrections is available.