Suppr超能文献

在法国国家辐射防护与核安全研究所(IRSN)的宽谱中子参考场中光子对周围剂量当量H(10)的贡献

Photon contribution to ambient dose equivalent H(10) in the wide-spectrum neutron reference fields of the IRSN.

作者信息

Langner F, Löb S, Nolte R, Gressier V, Asselineau B, Lacoste V, Lebreton L

机构信息

Physikalisch-Technische Bundesanstalt, Braunschweig, Germany.

出版信息

Radiat Prot Dosimetry. 2007;126(1-4):145-50. doi: 10.1093/rpd/ncm031. Epub 2007 May 25.

Abstract

The photon contribution to ambient dose equivalent in several wide-spectrum reference neutrons fields of the Institute for Radiological Protection and Nuclear Safety were measured using a Geiger-Müller counter. For the investigated fields, the ratio of photon to neutron ambient dose equivalent ranged between 0.03 and 0.20. The results show that the Geiger-Müller tube is a versatile instrument for dosimetry in mixed photon-neutron fields if sufficient information for the calculation of corrections is available.

摘要

使用盖革-米勒计数器测量了法国辐射防护与核安全研究所几个宽谱参考中子场中光子对环境剂量当量的贡献。对于所研究的场,光子与中子环境剂量当量之比在0.03至0.20之间。结果表明,如果有足够的信息用于计算修正值,盖革-米勒管是一种适用于混合光子-中子场剂量测定的通用仪器。

相似文献

7
The neutron component of two high-energy photon reference fields.两个高能光子参考射野的中子成分。
Radiat Prot Dosimetry. 2007;126(1-4):404-7. doi: 10.1093/rpd/ncm083. Epub 2007 Aug 2.
9
Improved measurements of neutron dose equivalent rates using a dual instrument technique.
Radiat Prot Dosimetry. 2006;122(1-4):393-5. doi: 10.1093/rpd/ncl489. Epub 2007 Feb 7.