Abstract

Five french (5F) catheters are being widely promoted for use in patients undergoing day case angiography including coronary angiography. Although there are theoretical advantages to this practice there are also potential disadvantages. We reviewed various performance parameters of four different brands of 5F coronary catheter and compared them with performance of the six (6F) and eight french (8F) coronary catheters used in routine coronary angiography. All the 5F catheters performed significantly less satisfactorily than the 6F and 8F catheters. 5F coronary catheters cannot be recommended for routine transfemoral coronary angiography.

摘要

5法式(5F)导管正被广泛推广用于接受日间血管造影(包括冠状动脉造影)的患者。尽管这种做法在理论上有优势,但也存在潜在的劣势。我们回顾了四种不同品牌5F冠状动脉导管的各种性能参数,并将它们与常规冠状动脉造影中使用的6法式(6F)和8法式(8F)冠状动脉导管的性能进行了比较。所有5F导管的表现明显不如6F和8F导管。不推荐将5F冠状动脉导管用于常规经股动脉冠状动脉造影。