Sandilos Panagiotis, Tatsis Elias, Vlachos Lampros, Karaiskos Pantelis, Georgiou Evangelos, Kipouros Panagiotis, Torrens Michael, Angelopoulos Angelos
Department of Radiology, Medical School, University of Athens, Areteion Hospital, Athens, Greece.
J Appl Clin Med Phys. 2006 Nov 28;7(4):13-21. doi: 10.1120/jacmp.v7i4.2273.
The polymer gel-Magnetic Resonance Imaging (MRI) dosimetry technique was employed to evaluate the mechanical and dose delivery accuracy in Leksell gamma knife stereotactic radiosurgery for the treatment of multiple targets. Two different polymer gel dosimeter formulations that have been reported in the literature were prepared in-house. A plan for the treatment of four brain metastases (targets) was generated. The plan involved the delivery of four, 8 mm collimator shots using different prescription isodose lines and different prescription doses for each target, keeping the maximum dose constant for all the targets. A sample of each gel formulation was irradiated using a custom made phantom with an experimental procedure capable of testing the increased nominal mechanical accuracy of stereotactic radiosurgery. The irradiated dosimeters were evaluated using a clinical 1.5 T Philips ACS NT MR imager. Result manipulation in 3-D allowed for the determination of the mechanical accuracy in the delivery of each shot through the comparison of measured versus planned shot center coordinates. Dose delivery accuracy was also evaluated by comparison of maximum dose values measured at the center of each shot as well as dose distribution measurements, with corresponding treatment planning calculations. Polymer gel dosimetry was found capable of verifying the complete chain of radiosurgery treatment in gamma knife applications involving the irradiation of multiple targets.
采用聚合物凝胶 - 磁共振成像(MRI)剂量测定技术,评估Leksell伽马刀立体定向放射外科治疗多个靶点时的机械精度和剂量传递准确性。制备了文献中报道的两种不同的聚合物凝胶剂量计配方。制定了一个治疗四个脑转移瘤(靶点)的计划。该计划包括使用不同的处方等剂量线和针对每个靶点的不同处方剂量,进行四次8毫米准直器照射,同时保持所有靶点的最大剂量恒定。使用定制的体模对每种凝胶配方的样本进行照射,该实验程序能够测试立体定向放射外科提高的标称机械精度。使用临床1.5T飞利浦ACS NT MR成像仪对辐照后的剂量计进行评估。通过三维结果处理,通过比较测量的与计划的照射中心坐标,确定每次照射传递中的机械精度。还通过比较每次照射中心测量的最大剂量值以及剂量分布测量结果与相应的治疗计划计算结果,评估剂量传递准确性。结果发现,聚合物凝胶剂量测定法能够验证伽马刀应用中涉及多个靶点照射的放射外科治疗的完整流程。