Tests of an electron monitor for routine quality control measurements of electron energies.

作者信息

Ramsay E B, Reinstein L E, Meek A G

机构信息

Department of Radiation Oncology, State University of New York, Stony Brook 11794-7028.

出版信息

Med Phys. 1991 Nov-Dec;18(6):1247-50. doi: 10.1118/1.596598.

Abstract

The depth dose for electrons is sensitive to energy and the AAPM Task Group 24 has recommended that tests be performed at monthly intervals to assure electron beam energy constancy by verifying the depth for the 80% dose to within +/- 3 mm. Typically, this is accomplished by using a two-depth dose ratio technique. Recently, a new device, the Geske monitor, has been introduced that is designed for verifying energy constancy in a single reading. The monitor consists of nine parallel plate detectors that alternate with 5-mm-thick absorbers made of an aluminum alloy. An evaluation of the clinical usefulness of this monitor for the electron beams available on a Varian Clinac 20 has been undertaken with respect to energy discrimination. Beam energy changes of 3 mm of the 80% dose give rise to measurable output changes ranging from 1.7% for 20-MeV electron beams to 15% for 6-MeV electron beams.