Tsujimura N, Yoshida T, Takada C, Momose T, Nunomiya T, Aoyama K
Nuclear Fuel Cycle Engineering Laboratories, Japan Atomic Energy Agency, 4-33, Tokai-mura, Ibaraki-ken 319-1194, Japan.
Radiat Prot Dosimetry. 2007;126(1-4):261-4. doi: 10.1093/rpd/ncm054. Epub 2007 Jun 2.
A new neutron-measuring instrument that is intended to measure a neutron personal dose equivalent, H(p)(10) was developed. This instrument is composed of two parts: (1) a conventional moderator-based neutron dose equivalent meter and (2) a neutron shield made of borated polyethylene, which covers a backward hemisphere to adjust the angular dependence. The whole design was determined on the basis of MCNP calculations so as to have response characteristics that would generally match both the energy and angular dependencies of H(p)(10). This new instrument will be a great help in assessing the reference values of neutron H(p)(10) during field testing of personal neutron dosemeters in workplaces and also in interpreting their readings.
一种旨在测量中子个人剂量当量H(p)(10)的新型中子测量仪器被研制出来。该仪器由两部分组成:(1) 传统的基于慢化剂的中子剂量当量仪;(2) 由硼酸聚乙烯制成的中子屏蔽体,它覆盖后半球以调节角依赖性。整个设计是基于MCNP计算确定的,以便具有通常能与H(p)(10)的能量和角依赖性相匹配的响应特性。这种新仪器将在评估工作场所个人中子剂量仪现场测试期间中子H(p)(10)的参考值以及解释其读数方面提供很大帮助。