Park H, Kim J, Choi K-O
Korea Research Institute of Standards and Science, Daejeon 305-600, South Korea.
Radiat Prot Dosimetry. 2007;126(1-4):159-62. doi: 10.1093/rpd/ncm034. Epub 2007 Jun 16.
For the purpose of radiation protection, the reference neutron field for calibration of neutron monitors was constructed using radioactive neutron sources-bare-(252)Cf, D(2)O-moderated (252)Cf and (241)Am-Be(alpha,n)-at Korea Research Institute of Standards and Science (KRISS). The well-specified neutron source with its emission rate and the anisotropy was installed at the centre of the neutron irradiation room, which is 6.6 x 7.6 x 6.3 m(3) in size. The neutron spectra of each source was measured using the Bonner sphere spectrometer (KRISS-BSS). Calculations using MCNP5 with realistic geometry and materials in the neutron irradiation room were performed. The calculations and measurements were found to be in good agreement, showing that the neutron calibration facility at KRISS is well established.
出于辐射防护目的,韩国标准科学研究院(KRISS)使用放射性中子源——裸(252)Cf、重水慢化(252)Cf和(241)Am-Be(α,n)构建了用于中子监测仪校准的参考中子场。将具有明确发射率和各向异性的特定中子源安装在尺寸为6.6×7.6×6.3立方米的中子辐照室中心。使用邦纳球谱仪(KRISS-BSS)测量了每个源的中子能谱。利用MCNP5对中子辐照室内实际的几何形状和材料进行了计算。计算结果与测量结果吻合良好,表明KRISS的中子校准设施已完善建立。