Józefowicz K, Golnik N, Tulik P, Zielczynski M
Laboratory of Radiation Protection Measurements, Institute of Atomic Energy, Otwock- Swierk, Poland.
Radiat Prot Dosimetry. 2007;126(1-4):134-7. doi: 10.1093/rpd/ncm029. Epub 2007 May 19.
The calibration laboratory, having standard neutron fields of radionuclide sources, should perform regular measurements of fields' parameters in order to check their stability and to get knowledge of any changes. Usually, accompanying gamma radiation is not of serious concern, but some personal dosemeters, old neutron dose equivalent meters with scintillation detectors and the dose meters of mixed radiation require the determination of this component. In the Laboratory of Radiation Protection Measurements in the Institute of Atomic Energy, Poland, the fields of radionuclide neutron sources (252)Cf, (241)Am-Be and (239)Pu-Be were examined for nearly 20 y. A number of detectors and methods have been applied for the determination of neutron ambient dose equivalent rate and for the determination of neutron and gamma dose components. This paper presents the recent results of measurements of gamma and neutron dose and dose equivalent, compared with the results accumulated in nearly 20 y.
校准实验室拥有放射性核素源的标准中子场,应定期对场的参数进行测量,以检查其稳定性并了解任何变化。通常,伴随的伽马辐射并不构成严重问题,但一些个人剂量计、带有闪烁探测器的旧中子剂量当量计以及混合辐射剂量计需要测定这一成分。在波兰原子能研究所辐射防护测量实验室,对放射性核素中子源(252)Cf、(241)Am-Be和(239)Pu-Be的场进行了近20年的检测。已应用多种探测器和方法来测定中子周围剂量当量率以及中子和伽马剂量成分。本文介绍了伽马和中子剂量及剂量当量测量的最新结果,并与近20年积累的结果进行了比较。