To investigate the tolerance of MR examinations in children and adolescents performed in a 1.5 Tesla MR scanner with an expanded bore diameter.
163 patients, ages 4 to 25, underwent MR examinations in a 1.5 Tesla MR scanner with an open design (MAGNETOM Espree, Siemens, Erlangen, Germany), characterized by a compact length of 125 cm and an expanded 70 cm bore diameter. MR imaging of the brain was carried out in most cases (78.5 %), followed by examinations of the spinal canal (9.8 %), the extremities (9.2 %) and the neck (2.5 %). The patients were divided into four age groups and the success rate, motion artifacts and diagnostic quality of the MR examinations were assessed using a 3-grade scale.
In 119 of 163 patients (73.0 %), MR examination was possible without any motion artifacts. With respect to the different age groups, 41.7 % of the 4 - 7-year-old children, 67.6 % of the 8 - 10-year-old children, 84.1 % of the 11 - 16-year-old children and 95.8 % of the patients older than 17 showed tolerance grade I without motion artifacts and excellent diagnostic image quality. In 39 of 163 children (23.9 %), the MR images showed moderate motion artifacts but had sufficient diagnostic quality. With regard to the different age groups, 52.8 % of the 4 - 7-year-old children, 26.5 % of the 8 - 10-year-old children, 15.9 % of the 11 - 16-year-old children and none of the patients older than 17 showed tolerance grade II with moderate motion artifacts and sufficient diagnostic image quality. In only 4 of 124 children < 10 years old and 1 child > 10 years old, the MR examination was not feasible and had to be repeated under sedation.
Pediatric MR imaging using a 1.5 Tesla MR scanner with an open design can be conducted in children and adolescents with excellent acceptance. The failure rate of 3.0 % of cases for pediatric MR imaging is comparable to that of a conventional low-field open MR scanner.
研究在孔径扩大的1.5特斯拉磁共振成像(MR)扫描仪中对儿童和青少年进行MR检查的耐受性。
163例年龄在4至25岁的患者在一台开放式设计的1.5特斯拉MR扫描仪(德国埃尔朗根西门子公司的MAGNETOM Espree)中接受了MR检查,该扫描仪的特点是长度紧凑,为125厘米,孔径扩大至70厘米。大多数情况下(78.5%)进行了脑部MR成像,其次是椎管检查(9.8%)、四肢检查(9.2%)和颈部检查(2.5%)。将患者分为四个年龄组,并使用三级量表评估MR检查的成功率、运动伪影和诊断质量。
163例患者中有119例(73.0%)可以进行无任何运动伪影的MR检查。对于不同年龄组,4至7岁儿童中有41.7%、8至10岁儿童中有67.6%、11至16岁儿童中有84.1%以及17岁以上患者中有95.8%表现为I级耐受性,无运动伪影且诊断图像质量优异。163例儿童中有39例(23.9%)的MR图像显示有中度运动伪影,但具有足够的诊断质量。对于不同年龄组,4至7岁儿童中有52.8%、8至10岁儿童中有26.5%、11至16岁儿童中有15.9%以及17岁以上患者中无一例表现为II级耐受性,有中度运动伪影且诊断图像质量足够。在124例10岁以下儿童中只有4例以及1例10岁以上儿童中,MR检查不可行,必须在镇静下重复进行。
使用具有开放式设计的1.5特斯拉MR扫描仪对儿童和青少年进行儿科MR成像时,患者接受度良好。儿科MR成像3.0%的失败率与传统低场开放式MR扫描仪相当。