Suppr超能文献

加强左心室辅助装置流出道移植物以防止扭结。

Reinforcement of left ventricular assist device outflow grafts to prevent kinking.

作者信息

Cohn William E, Gregoric Igor D, Frazier O H

机构信息

Center for Cardiac Support, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas 77225-0345, USA.

出版信息

Ann Thorac Surg. 2007 Jul;84(1):301-2. doi: 10.1016/j.athoracsur.2006.10.015.

Abstract

Because of their small size, rotary left ventricular assist devices (LVADs) can be implanted through several alternative approaches, including transdiaphragmatic LVAD insertion through a left subcostal incision with anastomosis of the outflow graft to the retroperitoneal supra-celiac aorta and a left thoracotomy approach with anastomosis to the descending thoracic aorta. More recently we have added a counter-incision to allow the outflow graft of a transdiaphragmatic LVAD to be tunneled through the right chest and anastomosed to the ascending aorta. However, constructing a tension-free, nonkinking lie of the outflow graft can be challenging. We have found that placing a 10-cm to 15-cm length of polytetrafluoroethylene graft coaxially over the LVAD outflow graft ensures a smooth, kink-free lie. Thus far 12 patients have undergone reinforcement of the LVAD outflow graft prior to graft-to-aorta anastomosis. In all cases, graft lie was facilitated and kinking was eliminated.

摘要

由于尺寸较小,旋转式左心室辅助装置(LVAD)可通过多种替代方法植入,包括经左肋下切口经膈肌插入LVAD,并将流出道移植物与腹膜后腹腔干上主动脉吻合,以及左胸廓切开术并与降主动脉吻合。最近,我们增加了一个辅助切口,使经膈肌LVAD的流出道移植物能够穿过右胸并与升主动脉吻合。然而,构建一个无张力、无扭结的流出道移植物走向可能具有挑战性。我们发现,在LVAD流出道移植物上同轴放置一段10厘米至15厘米长的聚四氟乙烯移植物,可确保其走向平滑、无扭结。到目前为止,已有12例患者在移植物与主动脉吻合前对LVAD流出道移植物进行了加固。在所有病例中,移植物走向均得到改善,扭结得以消除。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验