Bance Manohar, Campos Alfredo, Wong Lillian, Morris David P, van Wijhe Rene
E.A.R. Laboratory, Dalhousie University, Halifax, Nova Scotia, Canada.
Otolaryngol Head Neck Surg. 2007 Jul;137(1):70-3. doi: 10.1016/j.otohns.2007.01.019.
The transmission of vibrations from the tympanic membrane to the stapes footplate by an ossicular reconstruction prosthesis is affected by the size of the prosthesis head. We sought to determine if augmenting or reducing the head size of prosthesis had a systematic effect on transmission of vibrations to the stapes.
We conducted a fresh cadaveric temporal bone middle ear study.
The incus was replaced with a prosthesis using a tympanic membrane to stapes head (TASH)-type hydroxyapatite prosthesis in nine fresh cadaveric temporal bones. Three prosthesis head sizes were created: unaltered, reduced, and augmented. Stapes vibrations were measured with a laser Doppler vibrometer in response to acoustic frequency chirps at 90 dB SPL.
All three head size prostheses resulted in smaller stapes vibrations than the intact ear. There was no difference in the vibration transmission between the three different head sizes. All prostheses showed a vibration loss of 10 to 15 dB compared to the intact ear.
Within the range of sizes tested, prosthesis head size had little impact on vibration transmission to the stapes footplate.
听骨重建假体将振动从鼓膜传递至镫骨足板的过程会受到假体头部大小的影响。我们试图确定增大或减小假体头部尺寸是否会对向镫骨的振动传递产生系统性影响。
我们进行了一项新鲜尸体颞骨中耳研究。
在九个新鲜尸体颞骨中,使用鼓膜到镫骨头(TASH)型羟基磷灰石假体替换砧骨。创建了三种假体头部尺寸:未改变、减小和增大。使用激光多普勒振动计测量镫骨对90 dB SPL声频啁啾的振动响应。
与完整耳朵相比,所有三种头部尺寸的假体都导致镫骨振动较小。三种不同头部尺寸之间的振动传递没有差异。与完整耳朵相比,所有假体的振动损失均为10至15 dB。
在测试的尺寸范围内,假体头部尺寸对向镫骨足板的振动传递影响很小。