Suppr超能文献

新型可调节长度听骨重建假体在颞骨中的初步研究结果。

First results of a novel adjustable-length ossicular reconstruction prosthesis in temporal bones.

作者信息

Gottlieb Peter K, Li Xiping, Monfared Ashkan, Blevins Nikolas, Puria Sunil

机构信息

Department of Mechanical Engineering, Stanford University, Stanford, California, U.S.A.

Department of Otolaryngology-Head and Neck Surgery, Beijing Anzhen Hospital, Capital Medical University, Beijing, China.

出版信息

Laryngoscope. 2016 Nov;126(11):2559-2564. doi: 10.1002/lary.25901. Epub 2016 Mar 12.

Abstract

OBJECTIVES/HYPOTHESIS: The performance of an ossicular replacement prosthesis (ORP) is influenced by its alignment and appropriate tension between the tympanic membrane and the stapes footplate. A novel ORP with a flexible element that potentially allows for length adjustment in situ is presented and tested for acoustic performance.

STUDY DESIGN

Laser Doppler vibrometry in fresh human cadaveric temporal bones was used to test the acoustic performance of the adjustable ORP relative to standard prostheses used for ossiculoplasty.

METHODS

The three-dimensional (3D) velocity of the stapes posterior crus was measured in the 0.2- to 20-kHz range using a Polytec CLV-3D laser Doppler vibrometer. The middle ear cavity was accessed through a facial recess approach. After measuring the normal response, the incus was removed and stapes velocity was measured in the disarticulated case, then after insertion of the new prosthesis, a conventional prosthesis (Kurz BELL Dusseldorf type), and a sculpted autologous incus prosthesis in each temporal bone. The 3D stapes velocity transfer function (SVTF) was calculated for each case and compared.

RESULTS

The novel ORP design restored stapes velocity to within 6 dB (on average) of the intact response. No significant differences in 3D-SVTF were found between the new, conventional, or autologous ORPs.

CONCLUSIONS

The inclusion of an in situ adjustable element into the ORP design did not adversely affect its acoustic performance. The adjustable element may increase the ease of achieving optimal ORP placement, especially through a facial recess approach.

LEVEL OF EVIDENCE

NA Laryngoscope, 126:2559-2564, 2016.

摘要

目的/假设:听骨置换假体(ORP)的性能受其对线以及鼓膜与镫骨足板之间适当张力的影响。本文介绍了一种带有柔性元件的新型ORP,该元件可能允许原位长度调节,并对其声学性能进行了测试。

研究设计

在新鲜人类尸体颞骨中使用激光多普勒振动测量法来测试可调节ORP相对于用于听骨成形术的标准假体的声学性能。

方法

使用Polytec CLV-3D激光多普勒振动计在0.2至20 kHz范围内测量镫骨后脚的三维(3D)速度。通过面隐窝入路进入中耳腔。在测量正常反应后,去除砧骨并在关节离断的情况下测量镫骨速度,然后在每个颞骨中插入新假体、传统假体(库尔兹BELL杜塞尔多夫型)和雕刻的自体砧骨假体后进行测量。计算每种情况下的3D镫骨速度传递函数(SVTF)并进行比较。

结果

新型ORP设计将镫骨速度恢复到完整反应的平均6 dB以内。新型、传统或自体ORP之间在3D-SVTF方面未发现显著差异。

结论

在ORP设计中加入原位可调节元件不会对其声学性能产生不利影响。该可调节元件可能会增加实现最佳ORP放置的便利性,尤其是通过面隐窝入路。

证据级别

无 喉镜,126:2559 - 2564,2016年。