Fulton R E, Thompson H G
Defence Research and Development Canada-Suffield. Medicine Hat, Alberta, Canada.
J Immunoassay Immunochem. 2007;28(3):227-41. doi: 10.1080/15321810701454730.
Rapid analyte measurement platform (RAMPtrade mark) fluorogenic hand-held immunoassays (HHAs) were evaluated for inclusivity/sensitivity, exclusivity/specificity, sample matrix effects, ruggedness/stability, and reproducibility in detection of ricin (RCA(60)), a potential biological threat agent. The limit of detection of HHAs for RCA(60) was 14 ng/mL or approximately 140 pg/test. HHAs were inclusive in detection of ricin RCA(60), RCA(120), ricin A chain, and ricin B chain and exclusive in discrimination of RCA(60) from other toxins. None of the sample matrices tested affected assay performance. RAMPtrade mark ratios were tolerant of increases in sample volume, however, a decrease in sample volume of 25% resulted in significantly increased readings. RAMPtrade mark readings were highly reproducible lot-to-lot, cartridge-to-cartridge, day-to-day, and reader-to-reader.
对快速分析物测量平台(RAMP商标)荧光手持式免疫分析(HHA)进行了评估,以检测其在检测蓖麻毒素(RCA(60),一种潜在的生物威胁制剂)时的包容性/敏感性、排他性/特异性、样品基质效应、耐用性/稳定性和重现性。HHA对RCA(60)的检测限为14 ng/mL或约140 pg/测试。HHA在检测蓖麻毒素RCA(60)、RCA(120)、蓖麻毒素A链和蓖麻毒素B链时具有包容性,在区分RCA(60)与其他毒素时具有排他性。所测试的样品基质均未影响分析性能。RAMP商标比率可耐受样品体积的增加,然而,样品体积减少25%会导致读数显著增加。RAMP商标读数在批次间、试剂盒间、日间和仪器间具有高度重现性。