Chen Hsin-Yi, Huang Mei-Ling, Tsai Yi-Yu, Hung Por-Tying
Department of Ophthalmology, China Medical University Hospital, Taiching City, Taiwan.
Optom Vis Sci. 2007 Jul;84(7):640-6. doi: 10.1097/OPX.0b013e3180dc9a00.
To study the diagnostic ability of scanning laser polarimetry with variable corneal compensation (GDx VCC) for early glaucomatous (EG) eyes and glaucoma-suspect (GS) eyes in a Taiwan Chinese population.
This prospective cross-sectional study included 82 EG eyes (mean deviation, MD: -3.32 +/- 2.20 dB), 45 GS eyes (MD: -2.43 +/- 2.16 dB), and 62 normal eyes. Retinal nerve fiber layer thickness of each subject was measured using GDx VCC and Humphrey Field Analyzer visual field testing. Measured GDx VCC parameters were compared among groups. The area under the receiver operating characteristic (AROC) curve of each parameter was used to differentiate normal from EG eyes or GS eyes. The correlation between MD and each parameter was also evaluated.
For both normal versus EG and normal versus GS, the largest AROC values were for nerve fiber indicator, superior average thickness, and inferior average thickness. There was no significant correlation between MD and GDx-VCC-measured parameters either in EG or GS eyes.
GDx VCC shows only moderate ability to distinguish normal eyes from eyes with early glaucoma. However, its diagnostic role in eyes with suspicious discs and normal visual fields is uncertain at this moment in the Taiwan Chinese population. Further studies are needed to address this issue.
研究采用可变角膜补偿的扫描激光偏振仪(GDx VCC)对台湾华裔人群中早期青光眼(EG)眼和青光眼可疑(GS)眼的诊断能力。
这项前瞻性横断面研究纳入了82只EG眼(平均偏差,MD:-3.32±2.20 dB)、45只GS眼(MD:-2.43±2.16 dB)和62只正常眼。使用GDx VCC和Humphrey视野分析仪对每位受试者的视网膜神经纤维层厚度进行测量。对各组测量的GDx VCC参数进行比较。使用每个参数的受试者操作特征(AROC)曲线下面积来区分正常眼与EG眼或GS眼。还评估了MD与每个参数之间的相关性。
对于正常眼与EG眼以及正常眼与GS眼的比较,最大的AROC值出现在神经纤维指标、上方平均厚度和下方平均厚度上。在EG眼或GS眼中,MD与GDx-VCC测量的参数之间均无显著相关性。
GDx VCC在区分正常眼与早期青光眼眼方面仅表现出中等能力。然而,在台湾华裔人群中,其在视盘可疑且视野正常的眼中的诊断作用目前尚不确定。需要进一步研究来解决这个问题。