为首台Equinox-80型远距钴治疗机调试电动楔形板并在Eclipse 3D治疗计划系统中实施。

Commissioning of motorized wedge for the first equinox-80 telecobalt unit and implementation in the Eclipse 3D treatment planning system.

作者信息

Kinhikar R A, Sharma S, Upreti R, Tambe C M, Deshpande D D, Shrivastava S K, Dinshaw K A

机构信息

Department of Medical Physics, Tata Memorial Hospital, Parel, Mumbai, India.

出版信息

Australas Phys Eng Sci Med. 2007 Jun;30(2):127-34. doi: 10.1007/BF03178417.

Abstract

A new model of the telecobalt unit (TCU), Theratron Equinox-80, (MDS Nordion, Canada) equipped with a single 60 degree motorized wedge (MW), four universal wedges (UW) for 15 degrees, 30 degrees, 45 degrees and 60 degrees have been evaluated. MW was commissioned in Eclipse (Varian, Palo Alto, USA) 3D treatment planning system (TPS). The profiles and central axis depth doses (CADD) were measured with Wellhofer blue water phantom for MW and the measured data was commissioned in Eclipse. These profiles and CADD for MW were compared with UW in a homogeneous phantom generated in Eclipse for various field sizes. The dose was also calculated in the same phantom at 10 cm depth. For the particular MW angle and the respective open and MW beam weights, the dose was measured for a field size of 10 cm x 10 cm in a MEDTEC water phantom at 10 cm depth with a 0.13 cc thimble ion chamber (Scanditronix Wellhofer, Uppsala, Sweden) and a NE electrometer (Nuclear Enterprises, UK). Measured dose with ion chamber was compared with the TPS calculated dose. MW angle verification was also done on the Equinox for four angles (15 degrees, 30 degrees, 45 degrees and 60 degrees). The variation in measured and calculated dose at 10 cm depth was within 2%. The measured and the calculated wedge angles were in good agreement within 2 degrees. The motorized wedges were successfully commissioned in Eclipse for four wedge angles.

摘要

对一种新型远距离钴治疗机(TCU)——Theratron Equinox - 80(加拿大MDS Nordion公司生产)进行了评估,该治疗机配备了一个60度电动楔形板(MW)以及四个分别用于15度、30度、45度和60度的通用楔形板(UW)。MW在Eclipse(美国帕洛阿尔托瓦里安公司)三维治疗计划系统(TPS)中进行了调试。使用Wellhofer蓝水模体测量了MW的剂量分布和中心轴深度剂量(CADD),并将测量数据在Eclipse中进行了调试。在Eclipse中生成的均匀模体中,针对不同射野大小,将MW的这些剂量分布和CADD与UW进行了比较。还在同一模体中计算了10 cm深度处的剂量。对于特定的MW角度以及相应的开放野和MW射野权重,在MEDTEC水模体中,使用0.13 cc指形电离室(瑞典乌普萨拉Scanditronix Wellhofer公司)和NE静电计(英国核企业公司),在10 cm深度处测量了10 cm×10 cm射野大小的剂量。将电离室测量的剂量与TPS计算的剂量进行了比较。还在Equinox上对四个角度(15度、30度、45度和60度)进行了MW角度验证。10 cm深度处测量剂量与计算剂量的差异在2%以内。测量的楔形角与计算的楔形角在2度范围内吻合良好。电动楔形板在Eclipse中针对四个楔形角成功进行了调试。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验