Ding George X, Duggan Dennis M, Coffey Charles W, Deeley Matthew, Hallahan Dennis E, Cmelak Anthony, Malcolm Arnold
Department of Radiation Oncology, Vanderbilt University, Nashville, TN 37232-5671, USA.
Radiother Oncol. 2007 Oct;85(1):116-25. doi: 10.1016/j.radonc.2007.06.015. Epub 2007 Aug 20.
Changes in tumor size during the course of radiotherapy warrant performing adaptive radiotherapy (ART). This work investigates the feasibility and usefulness of acquiring on-board cone-beam CT (CBCT) for ART for patients with bulky head and neck tumors treated with IMRT and for prostate patients with potentially significant target position variations during the treatment course.
A phantom designed for CT quality assurance was used to compare the dosimetric and geometric accuracy between conventional CT and CBCT from a linear accelerator's on-board imager. Patient planning CT and CBCT images were acquired before treatment and at mid-course. The IMRT plans made on the CT were applied to the CBCT and dose-volume histograms were calculated.
In both phantom and patient studies, the dose-volume histograms (DVHs) based on CBCT images were in excellent agreement with DVHs based on planning CT images. Minimum, maximum and mean doses agreed very well. In a patient study, doses for targets and normal tissues from the same IMRT plans calculated on CBCT images agreed within 1-3% with those calculated on planning CT images.
CBCT images can be used to accurately predict dosimetric results. It is feasible to use CBCT to determine dosimetric consequences resulting from tumor shrinkage and patient geometry changes. An additional planning CT may be necessary to perform IMRT re-planning at present in order to accurately delineate tumor and organs. The CBCT has potential to become a very useful tool for on-line ART.
在放射治疗过程中肿瘤大小的变化需要进行自适应放射治疗(ART)。本研究探讨了对头颈部大体积肿瘤接受调强放疗(IMRT)的患者以及在治疗过程中靶区位置可能发生显著变化的前列腺癌患者,利用机载锥形束CT(CBCT)进行ART的可行性和实用性。
使用为CT质量保证设计的模体,比较来自直线加速器机载成像仪的传统CT和CBCT之间的剂量学和几何精度。在治疗前和疗程中期采集患者的计划CT和CBCT图像。将基于CT制定的IMRT计划应用于CBCT,并计算剂量体积直方图。
在模体和患者研究中,基于CBCT图像的剂量体积直方图(DVH)与基于计划CT图像的DVH高度一致。最小、最大和平均剂量的一致性非常好。在一项患者研究中,基于CBCT图像计算的同一IMRT计划对靶区和正常组织的剂量与基于计划CT图像计算的剂量相差1%-3%。
CBCT图像可用于准确预测剂量学结果。利用CBCT确定肿瘤缩小和患者几何形状变化所导致的剂量学后果是可行的。目前为了准确勾画肿瘤和器官,可能需要额外的计划CT来进行IMRT重新计划。CBCT有潜力成为在线ART的非常有用的工具。