Albert T J, Sharkey P F, Chao W, Hume E L, Rothman R H
Thomas Jefferson University, Philadelphia, PA.
J Arthroplasty. 1991;6 Suppl:S67-71. doi: 10.1016/s0883-5403(08)80058-7.
The authors have noted a change in the varus/valgus positioning of femoral components on anteroposterior (AP) radiographs that is dependent on limb rotation after total hip arthroplasty (THA). This effect, called pseudoposition, was demonstrated in six cadaver femurs implanted with an uncemented prosthesis that is noncanal-filling in the AP dimension. The distal tip of the prosthesis was positioned posteriorly in all specimens due to the bow of the femur. External rotation causes pseudovalgus positioning, and internal rotation causes pseudovarus positioning. Awareness of this phenomenon is important in the longitudinal analysis of femoral components after THA and emphasizes the need for standardized radiographic technique.
作者们注意到,全髋关节置换术(THA)后,前后位(AP)X线片上股骨组件的内翻/外翻定位会发生变化,这种变化取决于肢体的旋转。这种效应被称为假定位,在六具植入了非骨水泥假体的尸体股骨中得到了证实,该假体在AP维度上不填充髓腔。由于股骨的弓形,假体的远端在所有标本中均位于后方。外旋会导致假外翻定位,而内旋会导致假内翻定位。了解这一现象对于THA后股骨组件的纵向分析很重要,并强调了标准化放射学技术的必要性。