Polat Buelent, Wilbert Juergen, Baier Kurt, Flentje Michael, Guckenberger Matthias
Department of Radiation Oncology, Julius Maximilians University, Würzburg, Germany.
Strahlenther Onkol. 2007 Sep;183(9):506-11. doi: 10.1007/s00066-007-1747-5.
To investigate the magnitude and clinical relevance of relative motion/nonrigid setup errors in the head-and-neck (H&N) region.
Eleven patients with tumors in the H&N region were immobilized in thermoplastic head masks. Patient positioning was verified using a kilovoltage cone-beam CT (kv CBCT) prior to 100 treatment fractions. Five different regions of interest (ROIs) were selected for automatic image registration of planning CT and verification CBCT: (1) the whole volume covering planning CT and CBCT, (2) the skull, (3) the mandible, (4) C1-C3, and (5) C4-C6. Differences were calculated describing relative motion between the ROIs.
The 3-D patient setup error was 3.2 mm +/- 1.7 mm based on registration of the whole volume. No systematic relative motion (group mean errors <0.5 mm and <0.5 degrees ) between planning and treatment for any ROI was observed. Mobility was largest for the skull and the mandible relative to C4-C6 with 3-D displacements of 4.7 mm +/- 2.5 mm and 4.4 mm +/- 2.5 mm. Relative rotations were largest around the left-right axis (nodding) between C1-C3 and C4-C6 with maximum 11 degrees . No time trend of relative motion was observed. Margins for compensation of relative motion ranged between 5 mm and 10 mm.
The simplification of the patient as a rigid body was shown to result in significant errors due to relative motion in the H&N region. Margins for compensation of relative motion exceeded margins for compensation of patient positioning errors.
研究头颈部(H&N)区域相对运动/非刚性摆位误差的大小及其临床相关性。
11名头颈部肿瘤患者采用热塑性头面罩固定。在100次治疗分次前,使用千伏锥束CT(kv CBCT)验证患者体位。选择五个不同的感兴趣区域(ROI)用于计划CT和验证CBCT的自动图像配准:(1)覆盖计划CT和CBCT的整个体积,(2)颅骨,(3)下颌骨,(4)C1 - C3,以及(5)C4 - C6。计算描述ROI之间相对运动的差异。
基于整个体积的配准,三维患者摆位误差为3.2 mm±1.7 mm。未观察到任何ROI在计划和治疗之间存在系统性相对运动(组平均误差<0.5 mm且<0.5度)。相对于C4 - C6,颅骨和下颌骨的移动性最大,三维位移分别为4.7 mm±2.5 mm和4.4 mm±2.5 mm。C1 - C3和C4 - C6之间围绕左右轴(点头)的相对旋转最大,可达11度。未观察到相对运动的时间趋势。相对运动补偿边缘范围在5 mm至10 mm之间。
由于H&N区域的相对运动,将患者简化为刚体显示会导致显著误差。相对运动补偿边缘超过了患者体位误差补偿边缘。