Noble P A, Ashton E, Davidson C A, Albritton W L
Department of Applied Microbiology and Food Science, University of Saskatchewan, Saskatoon, Canada.
Appl Environ Microbiol. 1991 Nov;57(11):3287-91. doi: 10.1128/aem.57.11.3287-3291.1991.
Membrane filtration, spread plating, and pour plating are conventional methods used to determine the heterotrophic plate counts of water samples. Impedance methods were investigated as an alternative to conventional methods, since sample dilution is not required and the bacterial count can be estimated within 24 h. Comparisons of impedance signals obtained with different water samples revealed that capacitance produced faster detection times than conductance. Moreover, the correlation between heterotrophic plate count and detection time was highest (r = 0.966) when capacitance was used. Linear and quadratic regressions of heterotrophic plate count and impedance detection time were affected by incubation temperatures. Regressions between heterotrophic plate counts based on conventional methods and detection times of water samples incubated at less than or equal to 25 degrees C had R2 values of 0.878 to 0.933. However, regressions using detection times of water samples incubated at greater than or equal to 30 degrees C had lower R2 values, even though water samples produced faster detection times. Comparisons between broth-based versions of R2A medium and plate count agar revealed that the latter correlated highly with heterotrophic plate count, provided that water samples were incubated at 25 degrees C and impedance measurements were conducted with the capacitance signal (r = 0.966). When the linear regression of this relationship was tested with 100 water samples, the correlation between predicted and actual log10 CFU milliliter-1 was 0.869. These results indicate that impedance methods provide a suitable alternative to conventional methods.
膜过滤法、涂布平板法和倾注平板法是用于测定水样中异养平板计数的传统方法。由于不需要对样品进行稀释且可在24小时内估算出细菌数量,因此对阻抗法作为传统方法的替代方法进行了研究。对不同水样获得的阻抗信号进行比较后发现,电容法产生的检测时间比电导法更快。此外,使用电容法时,异养平板计数与检测时间之间的相关性最高(r = 0.966)。异养平板计数与阻抗检测时间的线性和二次回归受培养温度的影响。基于传统方法的异养平板计数与在小于或等于25摄氏度下培养的水样的检测时间之间的回归,R2值为0.878至0.933。然而,即使水样产生的检测时间更快,但使用在大于或等于30摄氏度下培养的水样的检测时间进行的回归,R2值较低。基于肉汤的R2A培养基和平板计数琼脂之间的比较表明,如果水样在25摄氏度下培养且使用电容信号进行阻抗测量,后者与异养平板计数高度相关(r = 0.966)。当用100个水样测试这种关系的线性回归时,预测的和实际的每毫升对数CFU之间的相关性为0.869。这些结果表明,阻抗法为传统方法提供了一种合适的替代方法。