柔性膀胱镜维修及成本的独立分析。

An independent analysis of flexible cystoscope repairs and cost.

作者信息

Canales Benjamin K, Gleason Joseph M, Hicks Nathan, Monga Manoj

机构信息

Department of Urologic Surgery, University of Minnesota, Minneapolis, Minnesota, USA.

出版信息

J Urol. 2007 Nov;178(5):2098-101; discussion 2102. doi: 10.1016/j.juro.2007.07.014. Epub 2007 Sep 17.

Abstract

PURPOSE

To establish the repair patterns of flexible cystoscopes we performed a systematic review of repairs performed at an independent endoscope repair company.

MATERIALS AND METHODS

Repair and cost data for 2000 to 2004 were obtained from an independent endoscope repair company. All cystoscopes underwent a detailed inspection using a preset protocol and the frequency of individual repair descriptions was extracted from a computer database.

RESULTS

A total of 1,150 repairs were performed on the Olympus CYF (3%), CYF-2 (21%), CYF-3 (56%) and CYF-4 (20%) flexible cystoscopes. Repairs to the distal tip (28% of all repairs) most commonly involved the bending rubber or outer polymer sheath. Repairs to the hand control segment (14% of all repairs) most commonly involved the ocular. Repairs for image (14% of all repairs) and deflection (9%) were most commonly performed on the fiberoptic bundle and wires, respectively. Repairs to the deflection apparatus comprised a significantly greater proportion of total repairs in the early CYF models than in the CYF-4 model (10% to 12% vs 3%, p <0.01, chi-square 10.73). Less than 1% of all cystoscopes needed more than 2 repairs.

CONCLUSIONS

Olympus cystoscopes require repair every 2 to 3 years and they appear to be durable. The distal deflection tip, specifically the outer bending rubber, is the most common site of flexible cystoscope damage. Although it appears that significant improvements have been made to the deflection apparatus, the number of repairs to other areas has not changed significantly with time. Improvements in form and function of the distal tip, the hand control segment and the fiberoptic bundles may enhance the durability of flexible cystoscopes.

摘要

目的

为确定软性膀胱镜的维修模式,我们对一家独立的内窥镜维修公司所进行的维修工作进行了系统评价。

材料与方法

获取了该独立内窥镜维修公司2000年至2004年的维修及成本数据。所有膀胱镜均按照预设方案进行详细检查,并从计算机数据库中提取各项维修描述的频率。

结果

共对奥林巴斯CYF(3%)、CYF - 2(21%)、CYF - 3(56%)和CYF - 4(20%)软性膀胱镜进行了1150次维修。远端尖端的维修(占所有维修的28%)最常涉及弯曲橡胶或外部聚合物护套。手动控制部分的维修(占所有维修的14%)最常涉及目镜。图像(占所有维修的14%)和偏转(9%)的维修最常分别针对光纤束和电线进行。在CYF早期型号中,偏转装置的维修占总维修的比例明显高于CYF - 4型号(10%至12%对3%,p<0.01,卡方值10.73)。所有膀胱镜中不到1%需要进行2次以上维修。

结论

奥林巴斯膀胱镜每2至3年需要维修一次,且似乎耐用。远端偏转尖端,特别是外部弯曲橡胶,是软性膀胱镜损坏最常见的部位。尽管偏转装置似乎有了显著改进,但随着时间推移,其他部位的维修次数并未明显改变。远端尖端、手动控制部分和光纤束在形式和功能上的改进可能会提高软性膀胱镜的耐用性。