Goertzen Andrew L, Suk Joon Young, Thompson Christopher J
Department of Radiology, University of Manitoba, Winnipeg, Manitoba, Canada.
J Nucl Med. 2007 Oct;48(10):1692-8. doi: 10.2967/jnumed.107.040584. Epub 2007 Sep 14.
Lutetium oxyorthosilicate (LSO)- or lutetium-yttrium oxyorthosilicate (LYSO)-based PET scanners have intrinsic radioactivity in the scintillator crystals due to the presence of (176)Lu, which decays by beta-emission followed by one or more prompt gamma-ray emissions. This leads to intrinsic true counts that can influence the image when scanning low levels of activity. An evaluation of the effects of this intrinsic activity for low levels of activity and different energy windows is performed on an LSO-based small-animal PET scanner.
Intrinsic count rate and sensitivity were measured for a range of lower-level discriminators (LLDs) ranging from 100 to 750 keV. The noise equivalent count rate (NECR) as a function of LLD for activity levels from 100 Bq to 100 kBq was estimated using a combination of measurement and previously published data for this scanner. Phantom imaging was performed using three (68)Ge sources of strength 55, 220, and 940 Bq and LLD levels of 250, 350, and 400 keV. The images were assessed using a contrast-to-noise ratio (CNR) analysis and by comparing the observed ratio of source activities to the true ratio value.
The intrinsic true count rate is reduced from 940 counts per second (cps) for a 250- to 750-keV energy window to <2 cps for a 400- to 750-keV window. There is a corresponding 2-fold drop in sensitivity for detected true events for external positron sources for these 2 energy windows. The NECR versus LLD curves showed a highly peaked shape, with the optimum LLD being approximately 425 keV. The phantom image results were dominated by the intrinsic true counts when an energy window of 250-750 keV was used. The intrinsic true counts were almost completely removed by raising the LLD to 400 keV. The CNR for each of the sources was higher for the narrow energy window and the 55 Bq could be easily visualized in images acquired with LLD levels of 350 and 400 keV but not when the 250-keV LLD was used. Images acquired with an LLD of 400 keV and reconstructed with 2-dimensional filtered backprojection were the most quantitatively accurate.
It is possible to visualize sources of <1 kBq in LSO-based animal PET systems by raising the LLD to 400 keV to exclude the majority of the counts due to the intrinsic activity present in the LSO.
基于硅酸镥(LSO)或硅酸镥钇(LYSO)的正电子发射断层扫描(PET)扫描仪,由于存在(176)Lu,闪烁晶体中存在固有放射性,(176)Lu通过β衰变,随后有一个或多个瞬发γ射线发射。这会导致固有真计数,在扫描低活度时会影响图像。在一台基于LSO的小动物PET扫描仪上,对这种固有活度在低活度和不同能量窗情况下的影响进行了评估。
对一系列下限甄别器(LLD),范围从100到750keV,测量了固有计数率和灵敏度。利用测量值和该扫描仪先前发表的数据相结合,估计了100Bq到100kBq活度水平下,噪声等效计数率(NECR)作为LLD的函数。使用三个强度分别为55、220和940Bq的(68)Ge源以及250、350和400keV的LLD水平进行体模成像。通过对比噪声比(CNR)分析以及比较观察到的源活度比与真实比值来评估图像。
对于250至750keV的能量窗,固有真计数率从每秒940计数(cps)降至400至750keV窗的<2cps。对于这两个能量窗,外部正电子源检测到的真事件灵敏度相应下降了2倍。NECR与LLD曲线呈高度峰值形状,最佳LLD约为425keV。当使用250 - 750keV能量窗时,体模图像结果主要受固有真计数影响。将LLD提高到400keV,固有真计数几乎完全消除。对于每个源,窄能量窗的CNR更高,在LLD水平为350和400keV采集的图像中,55Bq的源很容易可视化,但使用250keV LLD时则不行。使用400keV的LLD采集并通过二维滤波反投影重建的图像在定量方面最准确。
通过将LLD提高到400keV以排除由于LSO中存在固有活度产生的大部分计数,在基于LSO的动物PET系统中有可能可视化<1kBq的源。