Mischkowski Robert A, Ritter Larry, Neugebauer Jörg, Dreiseidler Timo, Keeve Erwin, Zöller Joachim E
Department of Oral and Maxillofacial Surgery, Umiversity of Cologne, Germany.
Quintessence Int. 2007 Oct;38(9):763-72.
To compare the diagnostic quality of 2-dimensional panoramic views generated from imaging data acquired by a newly developed cone beam machine with the diagnostic quality of conventional digital orthopantomograms (OPTs).
Thirty panoramic views reconstructed from digital volume tomograms (DVT) obtained by the pre-retail version of Galileos (Sirona Dental Systems), a newly developed compact cone beam device, were compared with 30 OPT images acquired by Orthophos XG Plus (Sirona Dental Systems). Diagnostic quality was assessed by 3 observers on criteria regarding detection of diagnosis-related findings, image quality, and visualization of anatomic structures in the maxillomandibular area. Wilcoxon rank sum test was applied for paired comparison.
Statistical analysis showed no significant difference in diagnostic quality between both imaging modes using a calculation model based on 10 weighted criteria (P = .629). There was especially no statistically significant difference in the assessment of diagnosis-related findings detection (P = .163). Image quality of DVT panoramic views was significantly lower than that of OPT. Except for mandibular canal and alveolar ridge gingiva, all evaluated anatomic structures were visualized better by DVT. Intraclass correlation coefficients were consistently higher in the DVT group for all criteria but 2, although the difference was not statistically significant (P = .374).
DVT panoramic views perform better in diagnosis of specific lesions, whereas OPTs provide a better image quality for a general overview of the maxillomandibular area. The diagnostic quality of both imaging modes is, however, equal.
比较一种新开发的锥形束机器采集的成像数据生成的二维全景视图与传统数字曲面断层片(OPT)的诊断质量。
将由新开发的紧凑型锥形束设备Galileos(西诺德牙科系统公司)零售前版本获取的数字容积断层扫描(DVT)重建的30幅全景视图与Orthophos XG Plus(西诺德牙科系统公司)采集的30幅OPT图像进行比较。3名观察者根据与诊断相关的发现的检测、图像质量以及颌面部区域解剖结构的可视化标准对诊断质量进行评估。采用Wilcoxon秩和检验进行配对比较。
统计分析表明,使用基于10个加权标准的计算模型,两种成像模式在诊断质量上无显著差异(P = 0.629)。在与诊断相关的发现检测评估方面尤其无统计学显著差异(P = 0.163)。DVT全景视图的图像质量显著低于OPT。除下颌管和牙槽嵴牙龈外,DVT对所有评估的解剖结构的可视化效果更好。除2项标准外,DVT组所有标准的组内相关系数始终较高,尽管差异无统计学意义(P = 0.374)。
DVT全景视图在特定病变的诊断中表现更好,而OPT为颌面部区域的总体概况提供了更好的图像质量。然而,两种成像模式的诊断质量相当。