Cheng Chee-Wai, Cho Sang Hyun, Taylor Michael, Das Indra J
Department of Radiation Oncology, Morristown Memorial Hospital, Morristown, New Jersey 07962, USA.
Med Phys. 2007 Aug;34(8):3149-57. doi: 10.1118/1.2750968.
In this study, zero-field percent depth dose (PDD) and tissue maximum ratio (TMR) for 6 MV x rays have been determined by extrapolation from dosimetric measurements over the field size range 1 x 1-10 x 10 cm2. The key to small field dosimetry is the selection of a proper dosimeter for the measurements, as well as the alignment of the detector with the central axis (CAX) of beam. The measured PDD results are compared with those obtained from Monte Carlo (MC) simulation to examine the consistency and integrity of the measured data from which the zero-field PDD is extrapolated. Of the six most commonly used dosimeters in the clinic, the stereotactic diode field detector (SFD), the PTW Pinpoint, and the Exradin A14 are the most consistent and produce results within 2% of each other over the entire field size range 1 x 1-40 x 40 cm2. Although the diamond detector has the smallest sensitive volume, it is the least stable and tends to disagree with all other dosimeters by more than 10%. The zero-field PDD data extrapolated from larger field measurements obtained with the SFD are in good agreement with the MC results. The extrapolated and MC data agree within 2.5% over the clinical depth range (dmax-30 cm), when the MC data for the zero field are derived from a 1 X 1 cm2 field simulation using a miniphantom (1 x 1 x 48 cm3). The agreement between the measured PDD and the MC data based on a full phantom (48 x 48 x 48 cm3) simulation is fairly good within 1% at shallow depths to approximately 5% at 30 cm. Our results seem to indicate that zero-field TMR can be accurately calculated from PDD measurements with a proper choice of detector and a careful alignment of detector axis with the CAX.
在本研究中,通过对1×1 - 10×10 cm²射野尺寸范围内的剂量学测量进行外推,确定了6 MV X射线的零野百分深度剂量(PDD)和组织最大剂量比(TMR)。小射野剂量学的关键在于选择合适的剂量仪进行测量,以及使探测器与射束中心轴(CAX)对齐。将测量得到的PDD结果与蒙特卡罗(MC)模拟结果进行比较,以检验用于外推零野PDD的测量数据的一致性和完整性。在临床中最常用的六种剂量仪中,立体定向二极管射野探测器(SFD)、PTW Pinpoint和Exradin A14最为一致,在整个1×1 - 40×40 cm²射野尺寸范围内,它们的结果相互之间相差在2%以内。尽管金刚石探测器的灵敏体积最小,但它最不稳定,与所有其他剂量仪的差异往往超过10%。用SFD在较大射野测量中得到的数据外推得到的零野PDD数据与MC结果吻合良好。当零野的MC数据来自使用小型模体(1×1×48 cm³)的1×1 cm²射野模拟时,在临床深度范围(dmax - 30 cm)内,外推数据与MC数据的吻合度在2.5%以内。基于完整模体(48×48×48 cm³)模拟得到的测量PDD与MC数据之间的吻合度在浅深度时相当好,在1%以内,在30 cm深度时约为5%。我们的结果似乎表明,通过正确选择探测器并使探测器轴与CAX仔细对齐,零野TMR可以从PDD测量中准确计算出来。