Failure of a dynamic hip screw (DHS) fixation leads to decreased mobility of the patient and frequently to a decrease in general health. The most common mode of failure of a DHS is cut out of the lag screw from the femoral head. The second most common mode of failure is lift-off of the plate from the femur. The aim of this laboratory-based experimental study was to determine whether a DHS secured to an osteoporotic femur with a locking screw plate would provide a stronger construct than the standard DHS plate.
The standard DHS design was compared to a DHS with fixed angle locking screws holding the DHS plate to the femur. Standard dynamic compression plates (DCP) and locking compression plates (LCP) were attached to synthetic, osteoporotic bone. A load was applied to replicate the forces occurring following the fixation of unstable, intertrochanteric hip fractures. A bracket on the proximal end of the plate replicated the lag screw in the femoral head. The constructs were cyclically loaded by a screw-driven material-testing machine and the number of cycles before failure occurred was determined.
The mean number of cycles to failure for the locking plate construct was 2.6 times greater than for the standard screw construct (285 versus 108 cycles, respectively p=0.016).
A dynamic hip screw with fixed angle locking screws would reduce the risk of DHS failure. A locking screw DHS would be particularly useful in patients with osteoporotic bone, and in patients with less stable fracture configurations.
动力髋螺钉(DHS)固定失败会导致患者活动能力下降,并常常使整体健康状况恶化。DHS最常见的失败模式是拉力螺钉从股骨头中穿出。第二常见的失败模式是钢板从股骨上抬起。这项基于实验室的实验研究的目的是确定用锁定螺钉钢板固定在骨质疏松性股骨上的DHS是否能提供比标准DHS钢板更强的结构。
将标准DHS设计与通过固定角度锁定螺钉将DHS钢板固定在股骨上的DHS进行比较。将标准动力加压钢板(DCP)和锁定加压钢板(LCP)附着在合成的骨质疏松性骨上。施加负荷以模拟不稳定型股骨转子间髋部骨折固定后产生的力。钢板近端的一个支架模拟股骨头中的拉力螺钉。通过螺旋驱动材料试验机对结构进行循环加载,并确定失效前的循环次数。
锁定钢板结构的平均失效循环次数比标准螺钉结构多2.6倍(分别为285次和108次循环,p = 0.016)。