Accuracy of ventriculostomy catheter placement using a head- and hand-tracked high-resolution virtual reality simulator with haptic feedback.
作者信息
Banerjee P Pat, Luciano Cristian J, Lemole G Michael, Charbel Fady T, Oh Michael Y
机构信息
Department of Mechanical and Industrial Engineering, College of Engineering, University of Illinois at Chicago 60607, USA.
出版信息
J Neurosurg. 2007 Sep;107(3):515-21. doi: 10.3171/JNS-07/09/0515.
OBJECT
The purpose of this study was to evaluate the accuracy of ventriculostomy catheter placement on a head- and hand-tracked high-resolution and high-performance virtual reality and haptic technology workstation.
METHODS
Seventy-eight fellows and residents performed simulated ventriculostomy catheter placement on an ImmersiveTouch system. The virtual catheter was placed into a virtual patient's head derived from a computed tomography data set. Participants were allowed one attempt each. The distance from the tip of the catheter to the Monro foramen was measured.
RESULTS
The mean distance (+/- standard deviation) from the final position of the catheter tip to the Monro foramen was 16.09 mm (+/- 7.85 mm).
CONCLUSIONS
The accuracy of virtual ventriculostomy catheter placement achieved by participants using the simulator is comparable to the accuracy reported in a recent retrospective evaluation of free-hand ventriculostomy placements in which the mean distance from the catheter tip to the Monro foramen was 16 mm (+/- 9.6 mm).