Suppr超能文献

移动化学探测器(AP2C+SP4E)作为战场医疗决策的辅助工具。

Mobile chemical detector (AP2C+SP4E) as an aid for medical decision making in the battlefield.

作者信息

Eisenkraft Arik, Markel Gal, Simovich Shirley, Layish Ido, Hoffman Azik, Finkelstein Arseny, Rotman Eran, Dushnitsky Tsvika, Krivoy Amir

机构信息

CBRN Medical Branch, Medical Corps, Army Headquarters, Israel Defense Forces, Ganei-Tikva, 55900, Israel.

出版信息

Mil Med. 2007 Sep;172(9):997-1001. doi: 10.7205/milmed.172.9.997.

Abstract

The combination of the AP2C unit with the SP4E kit composes a lightweight mobile detector of chemical warfare agents (CWA), such as nerve and mustard agents, with both vapor- and liquid-sampling capabilities. This apparatus was recently introduced into our military medical units as an aid for detection of CWA on casualties. Importantly, critical information regarding the applicability in the battlefield was absent. In view of the serious consequences that might follow a proclamation of CWA recognition in battlefield, a high false-positive rate positions the utilization of this apparatus as a medical decision tool in question. We have therefore conducted a field experiment to test the false-positive rate as well as analyze possible factors leading to false-positive readings with this device. The experiment was carried out before and after a 4-day army field exercise, using a standard AP2C device, a SP4E surface sampling kit, and a specially designed medical sampling kit for casualties, intended for medical teams. Soldiers were examined at rest, after mild exercise, and after 4 days in the field. The readings with AP2C alone were compared to the combination of AP2C and SP4E and to the medical sampling kit. Various body fluids served as negative controls. Remarkably, we found a false-positive rate of 57% at rest and after mild exercise, and an even higher rate of 64% after the 4-day field exercise with the AP2C detector alone, as compared to almost no false-positive readings with the combination of AP2C and SP4E. Strikingly, the medical sampling kit has yielded numerous false-positive readings, even in normal body fluids such as blood, urine, and saliva. We therefore see no place for using the medical sampling kit due to an unaccepted high rate of false-positive readings. Finally, we have designed an algorithm that uses the entire apparatus of AP2C and SP4E as a reliable validation tool for medical triage in the setting of exposure to nerve agents in the battlefield.

摘要

AP2C装置与SP4E套件相结合,构成了一种轻型化学战剂(CWA)移动探测器,可探测神经毒剂和芥子气等化学战剂,具备蒸汽和液体采样能力。该设备最近被引入我们的军事医疗单位,用于辅助检测伤员身上的化学战剂。重要的是,缺乏关于其在战场上适用性的关键信息。鉴于在战场上宣布识别化学战剂可能会带来严重后果,高假阳性率使得该设备作为医疗决策工具的实用性受到质疑。因此,我们进行了一项实地实验,以测试假阳性率,并分析导致该设备出现假阳性读数的可能因素。实验在为期4天的陆军野外演习前后进行,使用标准的AP2C设备、SP4E表面采样套件以及专门为医疗队设计的用于伤员的医疗采样套件。在士兵休息时、轻度运动后以及野外4天后对其进行检查。将单独使用AP2C的读数与AP2C和SP4E组合的读数以及医疗采样套件的读数进行比较。各种体液用作阴性对照。值得注意的是,我们发现单独使用AP2C探测器时,在休息和轻度运动后假阳性率为57%,在为期4天的野外演习后甚至高达64%,而AP2C和SP4E组合几乎没有假阳性读数。令人惊讶的是,即使在血液、尿液和唾液等正常体液中,医疗采样套件也产生了大量假阳性读数。因此,由于不可接受的高假阳性率,我们认为医疗采样套件没有使用价值。最后,我们设计了一种算法,将AP2C和SP4E的整个设备用作战场上接触神经毒剂情况下医疗分类的可靠验证工具。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验