O'Neill Adrian T, Monteiro-Riviere Nancy, Walker Glenn M
Dept. of Biomed. Eng., North Carolina Univ., Chapel Hill, NC, USA.
Conf Proc IEEE Eng Med Biol Soc. 2006;2006:2836-9. doi: 10.1109/IEMBS.2006.259270.
A novel microfluidic device is presented which creates a linear serial dilution of two input fluid streams. This platform facilitates higher productivity as a component of a high throughput cytotoxicity testing strategy. A modeling solution is presented to create custom linear dilution schemes. The featured device creates a serial dilution of two solutions in the range of 1:9 through 9:1 across nine discrete dilutions. It has been validated to create a highly linear progression of dilutions with an R2 value of 0.9993. The device functions equivalently over a wide range of flow rates. The standard deviation of dilution values averages 0.76% over six flow rates spanning 0.5 to 16 microl min(-1).