Hood Donald C, Bach Michael, Brigell Mitchell, Keating David, Kondo Mineo, Lyons Jonathan S, Palmowski-Wolfe Anja M
Departmentsof Psychology, Columbia University, New York, NY, USA.
Doc Ophthalmol. 2008 Jan;116(1):1-11. doi: 10.1007/s10633-007-9089-2. Epub 2007 Oct 31.
The clinical multifocal electroretinogram (mfERG) is an electrophysiological test of local retinal function. With this technique, many local ERG responses, typically 61 or 103, are recorded from the cone-driven retina under light-adapted conditions. This document specifies guidelines for performance of the test. It also provides detailed guidance on technical and practical issues, as well as on reporting test results. The main objective of the guidelines is to promote consistent quality of mfERG testing and reporting within and among centers. These 2007 guidelines, from the International Society for Clinical Electrophysiology of Vision (ISCEV: http://www.iscev.org ), replace the ISCEV guidelines for the mfERG published in 2003.
临床多焦视网膜电图(mfERG)是一种用于检测局部视网膜功能的电生理检查。通过该技术,在明适应条件下,从视锥细胞驱动的视网膜记录到许多局部视网膜电图反应,通常为61个或103个。本文档规定了该检查的操作指南。它还提供了有关技术和实际问题以及报告检查结果的详细指导。这些指南的主要目的是促进各中心内部和之间mfERG检查及报告质量的一致性。这些2007年的指南由国际临床视觉电生理学会(ISCEV:http://www.iscev.org )发布,取代了2003年发布的ISCEV mfERG指南。