Bach Michael, Brigell Mitchell G, Hawlina Marko, Holder Graham E, Johnson Mary A, McCulloch Daphne L, Meigen Thomas, Viswanathan Suresh
University Eye Hospital, Killianstr. 5, 79106, Freiburg, Germany.
Doc Ophthalmol. 2013 Feb;126(1):1-7. doi: 10.1007/s10633-012-9353-y. Epub 2012 Oct 17.
The pattern electroretinogram (PERG) is a retinal response evoked by a contrast-reversing pattern, usually a black and white checkerboard, which provides information about macular and retinal ganglion cell function. This document from the International Society for Clinical Electrophysiology of Vision (www.iscev.org) is a scheduled revision of the ISCEV PERG Standard, which updates and replaces the 2007 update and all earlier versions. The standard defines a single minimum stimulus and recording protocol for clinical PERG testing to assist practitioners in obtaining good quality responses and to facilitate inter-laboratory comparison. The present revision tightens stimulus specifications, expands on steady-state PERG recording, addresses visual stimulus display distinctions (CRT vs. LCD), and provides a more explicit definition of response components.
图形视网膜电图(PERG)是由对比度反转图形(通常是黑白棋盘格)诱发的视网膜反应,它提供有关黄斑和视网膜神经节细胞功能的信息。这份来自国际视觉临床电生理学会(www.iscev.org)的文件是ISCEV PERG标准的定期修订版,它更新并取代了2007年的更新版及所有早期版本。该标准定义了用于临床PERG测试的单一最小刺激和记录方案,以帮助从业者获得高质量的反应并促进实验室间的比较。本次修订收紧了刺激规范,扩展了稳态PERG记录,阐述了视觉刺激显示差异(阴极射线管与液晶显示器),并对反应成分给出了更明确 的定义