Suppr超能文献

一种将通用型动物正电子发射断层扫描(PET)扫描仪转换为高分辨率的PET插入式设备的可行性研究。

A feasibility study of a prototype PET insert device to convert a general-purpose animal PET scanner to higher resolution.

作者信息

Wu Heyu, Pal Debashish, O'Sullivan Joseph A, Tai Yuan-Chuan

机构信息

Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri 63110, USA.

出版信息

J Nucl Med. 2008 Jan;49(1):79-87. doi: 10.2967/jnumed.107.044149. Epub 2007 Dec 12.

Abstract

UNLABELLED

We developed a prototype system to evaluate the feasibility of using a PET insert device to achieve higher resolution from a general-purpose animal PET scanner.

METHODS

The system consists of a high-resolution PET detector, a computer-controlled rotation stage, and a custom mounting plate. The detector consists of a cerium-doped lutetium oxyorthosilicate array (12 x 12 crystals, 0.8 x 1.66 x 3.75 mm(3) each) directly coupled to a position-sensitive photomultiplier tube (PS-PMT). The detector signals were fed into the scanner electronics to establish coincidences between the 2 systems. The detector was mounted to a rotation stage that is attached to the scanner via the custom mounting plate after removing the transmission source holder. The rotation stage was concentric with the center of the scanner. The angular offset of the insert detector was calibrated via optimizing point-source images. In all imaging experiments, coincidence data were collected from 9 angles to provide 180 degrees sampling. A (22)Na point source was imaged at different offsets from the center to characterize the in-plane resolution of the insert system. A (68)Ge point source was stepped across the axial field of view to measure the sensitivity of the system. A 23.2-g mouse was injected with 38.5 MBq of (18)F-fluoride and imaged at 3 h after injection for 2 h.

RESULTS

The transverse image resolution of the PET insert device ranges from 1.1- to 1.4-mm full width at half maximum (FWHM) without correction for the point-source dimension. This corresponds to approximately 33% improvement over the resolution of the original scanner (1.7- to 1.8-mm FWHM) in 2 of the 3 directions. The sensitivity of the device is 0.064% at the center of the field, 46-fold lower than the sensitivity of an existing animal PET scanner. The mouse bone scan had improved image resolution using the PET insert device over that of the existing animal PET scanner alone.

CONCLUSION

We have demonstrated the feasibility of using a high-resolution insert device in an existing PET scanner to provide high-resolution PET. A PET insert device with more detector modules will improve sensitivity and may become an alternative to special-purpose PET systems for high-resolution PET.

摘要

未标注

我们开发了一个原型系统,以评估使用正电子发射断层扫描(PET)插入装置从通用动物PET扫描仪获得更高分辨率的可行性。

方法

该系统由高分辨率PET探测器、计算机控制的旋转台和定制安装板组成。探测器由铈掺杂的正硅酸镥阵列(12×12晶体,每个晶体尺寸为0.8×1.66×3.75 mm³)直接耦合到位置灵敏光电倍增管(PS-PMT)构成。探测器信号被馈入扫描仪电子设备,以建立两个系统之间的符合事件。在移除透射源支架后,探测器安装在通过定制安装板连接到扫描仪的旋转台上。旋转台与扫描仪的中心同心。通过优化点源图像校准插入探测器的角度偏移。在所有成像实验中,从9个角度收集符合数据以提供180度采样。对一个²²Na点源在距中心不同偏移处进行成像,以表征插入系统的平面内分辨率。一个⁶⁸Ge点源在轴向视野内移动,以测量系统的灵敏度。给一只23.2克的小鼠注射38.5 MBq的¹⁸F-氟化物,并在注射后3小时成像2小时。

结果

PET插入装置的横向图像分辨率在未对点源尺寸进行校正的情况下,半高宽(FWHM)为1.1至1.4毫米。这相当于在三个方向中的两个方向上,比原始扫描仪的分辨率(1.7至1.8毫米FWHM)提高了约33%。该装置在视野中心的灵敏度为0.064%,比现有动物PET扫描仪的灵敏度低46倍。使用PET插入装置的小鼠骨扫描图像分辨率比单独使用现有动物PET扫描仪时有所提高。

结论

我们已经证明了在现有PET扫描仪中使用高分辨率插入装置来提供高分辨率PET的可行性。具有更多探测器模块的PET插入装置将提高灵敏度,并可能成为用于高分辨率PET的专用PET系统的替代方案。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验