Omsland Tone K, Emaus Nina, Gjesdal Clara G, Falch Jan A, Tell Grethe S, Forsen Lisa, Berntsen Gro K R, Meyer Haakon E
Institute of General Practice and Community Medicine, University of Oslo, Oslo, Norway.
J Clin Densitom. 2008 Apr-Jun;11(2):276-82. doi: 10.1016/j.jocd.2007.10.001. Epub 2007 Dec 26.
The purpose of this study was to assess the agreement of in vivo hip scans on 3 densitometers (1 GE Lunar DPX-IQ and 2 GE Lunar Prodigy scanners) and to evaluate whether the European Spine Phantom (ESP) was able to reproduce the in vivo variability. Sixteen subjects had 3 repeated scans (with repositioning) on each densitometer, and the ESP was measured on each densitometer at least 40 times. Mean differences between hip scans on the Prodigy scanners were small and insignificant, and the in vivo results were not significantly different from the in vitro results. Bland and Altman plots showed no systematic differences between the Prodigy scanners over the range of bone mineral density (BMD). On the other hand, differences between Prodigy and DPX-IQ changed systematically over the range of BMD. The ESP did not fully reproduce the in vivo difference between Prodigy and DPX-IQ. In conclusion, the ESP is a valid substitute when assessing agreement between Prodigy scanners. However, when assessing agreement between different types of scanners, substitution of in vivo with in vitro measurements should be made with caution.
本研究的目的是评估3台骨密度仪(1台GE Lunar DPX-IQ和2台GE Lunar Prodigy扫描仪)对活体髋关节扫描结果的一致性,并评估欧洲脊柱体模(ESP)是否能够重现活体的变异性。16名受试者在每台骨密度仪上进行了3次重复扫描(重新定位),并且在每台骨密度仪上对ESP进行了至少40次测量。Prodigy扫描仪之间髋关节扫描的平均差异很小且无统计学意义,活体结果与体外结果无显著差异。Bland和Altman图显示,在骨密度(BMD)范围内,Prodigy扫描仪之间没有系统差异。另一方面,Prodigy和DPX-IQ之间的差异在BMD范围内呈系统性变化。ESP没有完全重现Prodigy和DPX-IQ之间的活体差异。总之,在评估Prodigy扫描仪之间的一致性时,ESP是一种有效的替代物。然而,在评估不同类型扫描仪之间的一致性时,用体外测量替代活体测量应谨慎进行。