A comparison of phantoms for cross-calibration of lumbar spine DXA.

作者信息

Pearson D, Cawte S A, Green D J

机构信息

Department of Medical Physics, Nothingham City Hospital NHS Trust, Nothingham, UK.

出版信息

Osteoporos Int. 2002 Dec;13(12):948-54. doi: 10.1007/s001980200132.

Abstract

The aim of this project was to compare three phantoms used for cross-calibration of dual-energy X-ray absorptiometers with an in vivo cross-calibration. The phantoms used were the Bona Fide Phantom (BFP), the European Spine Phantom (ESP) and the GE Lunar Aluminum Spine Phantom (ASP). The cross calibration was for L2-L4 lumbar spine bone mineral density (BMD) on a GE Lunar DPX-L and Hologic QDR 2000. The in vivo cross-calibration was obtained using 72 subjects (61 female, 11 male; mean age 49 years, range 14-84 years). The phantoms were measured 10 times without repositioning on both instruments. A further, long-term cross-calibration was obtained with the BFP over a 9 month period. The true linear relationship between the two instruments was calculated used a standardized principal components method. The mean residuals were calculated between each phantom cross-calibration line and the in vivo data to obtain a measure of the goodness of fit between the phantom cross-calibration and the in vivo data. There was no significant difference between the in vitro and in vivo cross-calibrations. The long-term BFP cross-calibration gave an in vitro cross-calibration that is closest to the in vivo cross-calibration in this group of subjects. When calculating Hologic QDR BMD from results on the GE Lunar DPX-L, the ASP underestimates Hologic QDR 2000 BMD by 4% at high BMD and overestimates by 4% at low BMD. The ESP cross-calibration overestimates Hologic QDR2000 BMD by 1% at high BMD and 4% at low BMD. The BFP performs best, overestimating Hologic QDR2000 BMD by between 1.2% and 1.8%, whilst the difference between the long-term BFP cross-calibration and the in vivo data is less than 1% over the range of BMD covered.

摘要

本项目的目的是将用于双能X线吸收仪交叉校准的三种体模与体内交叉校准进行比较。所使用的体模为正品体模(BFP)、欧洲脊柱体模(ESP)和GE Lunar铝制脊柱体模(ASP)。交叉校准针对的是GE Lunar DPX-L和Hologic QDR 2000上L2-L4腰椎的骨密度(BMD)。体内交叉校准是通过72名受试者(61名女性,11名男性;平均年龄49岁,范围14-84岁)获得的。在两台仪器上对体模进行了10次测量,且未重新定位。在9个月的时间里,使用BFP进行了进一步的长期交叉校准。使用标准化主成分法计算两台仪器之间的真实线性关系。计算每个体模交叉校准线与体内数据之间的平均残差,以获得体模交叉校准与体内数据之间拟合优度的度量。体外和体内交叉校准之间没有显著差异。长期的BFP交叉校准给出了在这组受试者中最接近体内交叉校准的体外交叉校准。当根据GE Lunar DPX-L的结果计算Hologic QDR BMD时,在高骨密度时ASP低估Hologic QDR 2000 BMD 4%,在低骨密度时高估4%。ESP交叉校准在高骨密度时高估Hologic QDR2000 BMD 1%,在低骨密度时高估4%。BFP表现最佳,高估Hologic QDR2000 BMD 1.2%至1.8%,而在涵盖的骨密度范围内,长期BFP交叉校准与体内数据之间的差异小于1%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验