Krivan Lubomír, Kozák Milan, Vlasínová Jitka, Sepsi Milan
Department of Cardiology, University Hospital Brno, Jihlavská 20, 625 00 Brno, Czech Republic.
Pacing Clin Electrophysiol. 2008 Jan;31(1):3-6. doi: 10.1111/j.1540-8159.2007.00917.x.
The authors present two cases of patients with perforation of the right ventricular wall by the implantable cardioverter defibrillator (ICD) lead. The complication was resolved by cardiosurgical revision and epicardial leads stitched onto the diaphragmatic wall of the heart. The perforation was identified by electrical parameter changes of the leads, echocardiography, and computed tomography. Both patients had satisfactory values of electrical parameters and ICD function with epicardial leads. The importance of regular follow-up and a check of the lead parameters are emphasized.
作者报告了两例植入式心脏复律除颤器(ICD)导线导致右心室壁穿孔的病例。通过心脏外科手术修复以及将心外膜导线缝合到心脏膈肌壁上,该并发症得以解决。通过导线的电参数变化、超声心动图和计算机断层扫描确定了穿孔情况。两名患者的心外膜导线电参数和ICD功能值均令人满意。强调了定期随访和检查导线参数的重要性。