Suppr超能文献

将Centrimag Levitronix泵与柏林心脏Excor插管相连;一种桥接至桥接的新方法。

Connecting the Centrimag Levitronix pump to Berlin Heart Excor cannulae; a new approach to bridge to bridge.

作者信息

Maat Alexander P W M, van Thiel Robert J, Dalinghaus Michiel, Bogers Ad J J C

机构信息

Department of Cardiothoracic Surgery, Erasmus Medical Center, Gravendijkwal 230 3015 CE Rotterdam, The Netherlands.

出版信息

J Heart Lung Transplant. 2008 Jan;27(1):112-5. doi: 10.1016/j.healun.2007.10.010.

Abstract

BACKGROUND

An increasing number of children are requiring circulatory support. Hospitals offering pediatric Ventricular Assist Device (VAD) should have devices of different sizes available to cover the full range of patient sizes incurring considerable expense. As in adults, post-operative bleeding often complicates VAD implantation. The use of a Levitronix Centrimag centrifugal pump, connected to Berlin Heart Excor cannulae, seems an attractive and logic combination, both in terms of patient safety and of hospital economics.

METHODS

We describe 3 children with therapy resistant cardiac failure who underwent extracorporeal membrane oxygenation (ECMO) as a rescue before proceeding to placement of the Berlin Heart Excor paracorporeal assist device. The Levitronix Centrimag pump was used as an intermediate device to allow the patients to be stabilised. Therefore, only Berlin Heart cannulae of different sizes have to be readily available; if successful stabilization can be achieved, the Berlin Heart Excor ventricles and the drive unit can then be ordered to replace the Levitronix pump.

RESULTS

Two patients were successfully stabilised with the Levitronix pump and were switched to the definitive Berlin Heart Excor ventricles after 6 days of support. The third child succumbed due to intractable pulmonary hemorrhage in severely damaged lungs. No device related complications, especially no thrombo-embolic events, occurred during Levitronix support.

CONCLUSION

The Levitronix Centrimag pump was easy to handle and gave effective circulatory support, the patients were only switched to the Berlin Heart Excor system after stabilization. In patients with a high risk of failure, it is a relatively cheap but safe and effective support system.

摘要

背景

需要循环支持的儿童数量日益增加。提供小儿心室辅助装置(VAD)的医院应备有不同尺寸的装置,以覆盖所有患者尺寸范围,这会产生相当高的费用。与成人情况一样,术后出血常常使VAD植入复杂化。连接到柏林心脏体外循环插管的Levitronix Centrimag离心泵的使用,在患者安全和医院经济方面似乎都是一种有吸引力且合理的组合。

方法

我们描述了3例治疗抵抗性心力衰竭儿童,在植入柏林心脏体外循环辅助装置之前,作为抢救措施接受了体外膜肺氧合(ECMO)治疗。Levitronix Centrimag泵用作中间装置以使患者稳定。因此,只需随时备有不同尺寸的柏林心脏插管;如果能成功实现稳定,随后可订购柏林心脏体外循环心室和驱动单元来替换Levitronix泵。

结果

2例患者通过Levitronix泵成功稳定,并在支持6天后切换至最终的柏林心脏体外循环心室。第三个孩子因严重受损肺部的顽固性肺出血而死亡。在Levitronix支持期间未发生与装置相关的并发症,尤其是未发生血栓栓塞事件。

结论

Levitronix Centrimag泵易于操作并能提供有效的循环支持,患者仅在稳定后才切换至柏林心脏体外循环系统。对于失败风险高的患者,它是一种相对便宜但安全有效的支持系统。