Moehrlen Ueli, Ziegler Urs, Weiss Markus
Department of Pediatric Surgery, University Children's Hospital Zurich, Steinwiesstrasse 75, CH-8032 Zurich, Switzerland.
Paediatr Anaesth. 2008 Mar;18(3):240-4. doi: 10.1111/j.1460-9592.2007.02386.x.
Technical and conceptual shortcomings of cuffed pediatric tracheal tubes have been pointed out in the past, all representing a potential risk for mucosal damage of the pediatric airway. In this study, cuffed pediatric tracheal tubes from different manufacturers were investigated with regard to the cuff shoulder design using scanning electron microscopy (SEM).
Cuffed pediatric tracheal tubes with the smallest internal diameter available were ordered from five manufacturers. SEM scans were performed from the tube shaft - cuff shoulder - tube cuff assembly and from the cuff shoulder alone. Each investigation was repeated in four samples of each tube brand. In addition, thickness of cuff membrane was measured.
Major differences were found among the cuff shoulders investigated ranging from smooth transitions, softened borders to very sharp steps. There was also considerable difference in the thickness among the cuff membranes between the different tube brands ranging from 18 microm in polyurethane cuffs up to 90 microm in polyvinylchloride cuffs.
Because the gap between the outer tracheal tube and the internal mucosal layer of the pediatric airway is often very narrow, some of the observed irregularities have the potential to injure the airway mucosa in pediatric patients, particularly during long-term use and firm fit of the tracheal tube in the airway. As cuffed tubes are increasingly used in small children, the observed irregularities in some tracheal tube brands may need to be improved and their safety should be proved by the manufacturers.
过去已指出小儿带套囊气管导管存在技术和概念上的缺陷,所有这些都对小儿气道黏膜造成潜在损伤风险。在本研究中,使用扫描电子显微镜(SEM)对不同制造商生产的小儿带套囊气管导管的套囊肩部设计进行了研究。
从五个制造商订购了内径最小的小儿带套囊气管导管。对管身 - 套囊肩部 - 管套组件以及单独的套囊肩部进行SEM扫描。每个品牌的每个导管在四个样本中重复进行每次研究。此外,测量了套囊膜的厚度。
在所研究的套囊肩部中发现了主要差异,范围从平滑过渡、边界软化到非常尖锐的台阶。不同品牌导管的套囊膜厚度也存在相当大的差异,聚氨酯套囊的厚度为18微米,聚氯乙烯套囊的厚度高达90微米。
由于小儿气道的气管导管外径与内部黏膜层之间的间隙通常非常狭窄,一些观察到的不规则情况有可能损伤小儿患者的气道黏膜,特别是在气管导管长期使用且在气道中固定良好的情况下。随着带套囊导管在小儿中的使用越来越多,一些气管导管品牌中观察到的不规则情况可能需要改进,制造商应证明其安全性。