[使用基于荧光透视的光电导航系统测定股骨前倾角:精度分析]

[Determining the femoral antetorsion angle with a fluoroscopy-based optoelectronic navigation system: a precision analysis].

作者信息

Keil C, von Recum J, Nolte L-P, Wentzensen A, Grützner P A

机构信息

Berufsgenossenschaftliche Unfallklinik Ludwigshafen, Klinik für Unfall- und Wiederherstellungschirurgie, Ludwig-Guttmann-Str. 13, 67071 Ludwigshafen, Deutschland.

出版信息

Unfallchirurg. 2008 Feb;111(2):126-31. doi: 10.1007/s00113-008-1406-z.

Abstract

According to the literature, differences in torsion of 15 degrees and more develop in 20-30% of cases after intramedullary nailing of femoral shaft fractures. A computer-assisted method makes it possible to determine the antetorsion angle during surgery. In this experimental study, the precision of the measurements obtained with the navigation system were checked with a femur model and compared with a CT reference method. The measurements are carried out on a femur model that is equipped with a rotation device in the middle of the shaft. Nine reproducible angles can be set. Two investigators each conduct the measurements of the antetorsion angle ten times. A comparison is drawn between the absolute values of the antetorsion angle measured and the difference values of the adjoining positions. When comparing the absolute values of the navigation and reference systems, the mean deviations of both methods are around 1 degrees (0.35; 1.75) and comparing the differences 0.5 degrees (-0.2; 1.17). The maximum deviation of the absolute values of the CT reference method amounts to 6.4 degrees . Under experimental conditions, measurement of the femoral antetorsion angle proved to be sufficiently precise for clinical specifications in comparison to a CT reference method.

摘要

根据文献报道,股骨干骨折髓内钉固定术后20%-30%的病例会出现15度及以上的扭转差异。计算机辅助方法能够在手术过程中确定前倾角。在本实验研究中,使用股骨模型检查导航系统测量的精度,并与CT参考方法进行比较。测量在一个股骨干中部装有旋转装置的股骨模型上进行。可以设置九个可重复的角度。两名研究人员分别对前倾角进行十次测量。对测量的前倾角绝对值与相邻位置的差值进行比较。比较导航系统和参考系统的绝对值时,两种方法的平均偏差约为1度(0.35;1.75),比较差值时为0.5度(-0.2;1.17)。CT参考方法绝对值的最大偏差为6.4度。在实验条件下,与CT参考方法相比,股骨前倾角的测量对于临床规范来说足够精确。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验